20
―――――――――――――――――――――――――――
Chapter 2
Measurement Procedure
――――――――――――――――――――――――
I/
key
U/
key
AUTO
range
150 V
range
300 V
range
600 V
range
AUTO
range
20 A
range
200 A
range
1000 A
range
Table 1.
Range Composition for Single-Phase Power Measurement
U
Voltage range
I
150.0 V
300.0 V
600 V
Current
range
20.00 A
3.000 k
6.000 k
12.00 k
200.0 A
30.00 k
60.00 k
120.0 k
1000 A
150.0 k
300.0 k
600.0 k
Unit
[W] or [VA] or [var]
2.3 Range Setup
1. Press the
RANGE
key. The voltage range then
appears on Display 2 and current range on Display
3. In this condition, Display 2 and Display 3
should be blinking.
2. To change the voltage range, press the
U/
key.
To change the current range, press the
I/
key.
The power range varies with the combination of
voltage and current ranges as listed in Tables 1 and
2.