82
Volara™ System, Model PVL1 User Manual (196286 REV C)
Cleaning and Disinfecting
•
Warning
—Clean the patient circuit between each therapy
session for the same patient.
•
Warning
—If the circuit and/or bio-filter are damaged or visibly
soiled, replace them. See “Replacement Parts/Kits” on page 78.
NOTE:
Each patient circuit is for use by a single patient and intended for 30 days
of treatment or a maximum of 90 treatment sessions.
These cleaning instructions are for multiple use of the circuit by a single
patient. The circuit has been tested for compatibility with mild detergent
and warm water for cleaning. The facility should follow their internal
infection control policies on cleaning the nebulizer and circuit. If a policy
is not in place, the following is a guideline for proper cleaning of the
circuit:
1.
Disconnect the circuit from the control unit.
2.
If a nebulizer is used in the therapy, disconnect the nebulizer from the
handset.
a.
Disconnect the nebulizer tubing from the medicine cup.
b.
Remove the lid from the medication cup.
c.
Discard any unused medication in accordance with facility
protocol.
3.
Disconnect the mouthpiece, face mask, adapters, handset, breathing
hose, and bio-filter.
4.
Examine the inside of the bio-filter for any damage and/or soilage. If
there is soilage or damage, replace the bio-filter immediately. For
examples of damaged bio-filters, see page 84.
5.
Wipe down the outside of the bio-filter with a compatible bacterial
cleaning solution. Do not use excessive liquid or harsh cleansers. See
“Compatible cleaning and disinfecting solutions” on page 80.
6.
Wash the mouthpiece or face mask, handset, and adapters with mild
detergent and warm water.
7.
Rinse the nebulizer cup and components with sterile water.
8.
Let the handset, nebulizer cup, mouthpiece, and adapters air dry or
dry them with a lint free cloth or paper towel.
9.
Wipe down the outside of the circuit tubing with an approved
alcohol-based cleaner.
10. After all parts are dry, assemble the circuit, and put it in a breathable
bag with the patient’s name on it for use later.
196654__C_2019-Apr-29_Cleaned