Volara™ System, Model PVL1 User Manual (196286 REV C)
Troubleshooting
019
1.
Make sure only the Hill-Rom approved
battery (194566S) is used.
2.
Press
return
after you read the on-screen
message.
3.
Make sure the battery is correctly installed
in the control unit. If necessary, remove the
battery and install it. Remove the battery
and install it. See “Install the Battery” on
page 67.
022
1.
Make sure the correct patient circuit is set
up for the required therapy. See page 16.
2.
Make sure a tight seal is maintained on the
face mask or mouthpiece.
3.
Examine the connections between the
patient circuit components, tubings, and
the control unit for any source of leakage. If
necessary, disassemble the patient circuit
and assemble it again. If you see cracks on
the patient circuit components, replace the
parts. See “Replacement Parts/Kits” on page
78.
4.
Press
return
after you read the on-screen
message.
023
1.
Make sure only the Hill-Rom approved bio-
filter (206680) is used. For instructions on
how to connect the filter correctly, see page
16.
2.
Press
return
after you read the on-screen
message.
Error Code
Do this:
196654__C_2019-Apr-29_Cleaned