ID3000 Series Smart Code Reader User Manual
42
4. (Optional) Go to Image Settings, and adjust parameters like exposure time, gain, Gamma, light
source, etc. if the code score is low.
Note
If the code score is still low after adjusting, and the code may have poor printing quality.
8.3 Signal Input Settings
In the signal input module, you can set the trigger related parameters. You can enable trigger
mode to let the acquisition of image data occur only when the trigger source is generated.
8.3.1 Set Trigger Mode
The device has 2 types of trigger mode: Internal trigger mode and external trigger mode.
Internal Trigger Mode
In this mode, the device acquires images via its internal signals.
External Trigger Mode
In this mode, the device acquires images via external signals like software signal and hardware
signal. The trigger source of external trigger mode includes software, physical lines, counter,
TCP, UDP, and serial.
8.3.2 Enable Internal Trigger Mode
In the internal trigger mode, the device acquires images via its internal signals. You have 2
methods to enable the internal trigger mode:
●
Click I/O Control Settings → Input → Trigger Mode, and select Off as Trigger Mode.
●
In the live view page, click
to enable the internal trigger mode.
8.3.3 Enable External Trigger Mode
In the external trigger mode, the device acquires images via external signals like software signal
and hardware signal. You have 2 methods to enable the external trigger mode:
●
Click I/O Control Settings → Input → Trigger Mode, and select On as Trigger Mode.
●
In the live view page, click
to enable the external trigger mode.
Set and Execute Software Trigger Mode
In software trigger, the software sends trigger signal to the device via I/O interface to acquire