54
Measurement functions
Temperature and humidity for guaranteed accuracy:
23°C±5°C (73.4°F ±8.5°F) and 90% RH or lower
Insulation Resistance Measurement
Rated mea-
surement volt-
age (DC)
50 V
125 V
250 V
500 V
1000 V
Effective maxi-
mum display
value
100 M
:
250 M
:
500 M
:
2000 M
:
4000 M
:
Effective medi-
um value
2 M
:
5 M
:
10 M
:
50 M
:
100 M
:
1st effective
measuring
range [M
:
]
0.200 to
10.00
0.200 to
25.0
0.200 to
50.0
0.200 to
500
0.200 to
1000
Accuracy
(Limit deviation
tolerance)
±4%rdg.
2nd effective
measuring
range [M
:
]
10.1 to
100.0
25.1 to
250
50.1 to
500
501 to
2000
1010 to
4000
Accuracy
(Limit deviation
tolerance)
±8%rdg.
Other measur-
ing range [M
:
]
0 to 0.199
Accuracy
(Limit deviation
tolerance)
±2%rdg.±6dgt.
www.
.com
1.800.561.8187