2.3 Inspecting the Instrument When Using the L9788-
⚝ᒁ
4
3
2
1
7
6
5
10
9
8
2.3
Inspecting the Instrument When
Using the L9788-10 Test Lead with
Remote Switch (Red)
To prevent an electric shock accident, confirm that the
white portion (insulation layer) inside the cable is not
exposed. If a color inside the cable is exposed, do not
use the cable.
1.
Confirm that the power of the Insulation Resistance
Tester is turned off.
2.
Plug the test lead securely into the LINE terminal of the
Insulation Resistance Tester.
3.
Short circuit the tips of the test leads.
4.
Use the rotary selector to select Insulation Resistance.
5.
Upon pressing the
MEASURE
key on the L9788-10,
verify that the L9788-10’s
MEASURE
key turns red when
the instrument’s live wire warning is displayed and that
the display shows 0 M
:
.
6.
Press
of the instrument to confirm that the light at the
tip of the L9788-10 comes on.
No
Yes