44
2.9 PV
:
Measurement
To avoid electric shock, short circuits and damage to the
instrument, observe the following precautions:
• Measurements with P-N shorted should use an
insulation resistance range other than PV
:
range.
• During hours such as in the nighttime while a solar
panel is not generating power, make measurements
with P-N shorted.
• Do not make any measurement if any of the bypass
diodes of the solar panel has failed. Doing so may
damage the solar panel.