![HIKOKI IR4051 Скачать руководство пользователя страница 61](http://html.mh-extra.com/html/hikoki/ir4051/ir4051_instruction-manual_2131756061.webp)
59
⚝ᒁ
4
3
2
1
7
6
5
10
9
8
*For only IR4053, the accuracy is guaranteed up to 1000 V.
Voltage Measurement
AC/DC
automatic
detection range
AC detected at 30 V or greater (50 Hz/60 Hz)
Any ripple current with superimposed current compo-
nent at 30 V or higher is identified as alternate current.
Effect of
temperature
Measurement accuracy per 1°C × 0.1
(Applicable for operational temperature range exclud-
ing 18°C to 28°C)
Overload
protection
750 V AC (10 s), 750 V DC (10 s)
1200 V DC (10 s) for IR4053 only
Display update
interval
Within 1 s
Response time
Within 1.2 s
(when input voltage is raised from 0 V to 600 V)
Input
resistance
100 k
:
or more (50 Hz/60 Hz)
Frequency
range
50 Hz/60 Hz
Range configuration
Display range
(Auto range)
Maximum
display value
Resolution
Accuracy
420 V
(Min. display
value: 30.0 V)
420.0 V
0.1 V
±2.3%rdg.±8dgt
(Accuracy of range
over 600 V is not
guaranteed)
600 V
750 V
1 V
Input
resistance
100 k
:
or more
Range configuration
Display range
(Auto range)
Maximum
display value
Resolution
Accuracy
4.2 V
4.200 V
0.001 V
±1.3%rdg.±4dgt
(Accuracy of range
over 600 V* is not
guaranteed)
42 V
42.00 V
0.01 V
420 V
420.0 V
0.1 V
IR4051:
600 V
IR4052:
600 V
IR4053:
1000 V
IR4051: 750 V
IR4052: 750 V
IR4053: 1100 V
1 V
A
C
Vo
lta
ge Measure
m
ent
DC V
o
lt
ag
e
Me
asureme
n
t
www.
.com
1.800.561.8187