32
T560MANK5.DOCX
5.
Powerup States and Saved Setups
Users may program the T560 as desired and then use the
SAve
command to copy all
setups to nonvolatile flash memory. That saved setup may be recalled at any time via the
REcall
command. The saved setup is also recalled and installed at powerup, allowing the
T560 to resume operation without any serial commands.
The powerup sequence takes about 4 seconds. During this time, channel outputs are
electrically low and terminations are high-Z.
If the saved configuration programs channels to be inverted polarity, those outputs will
transition from low to high when the initial powerup sequence is over.
Содержание T560
Страница 1: ...T560 DIGITAL DELAY GENERATOR Technical Manual February 13 2019 ...
Страница 15: ...15 T560MANK5 DOCX Figure 3 4 Output rising edge 100 ps delay steps ...
Страница 16: ...16 T560MANK5 DOCX Figure 3 5 Trigger and output pulses widths 4ns 3 ns 2 ns and 1 5 ns ...
Страница 35: ...35 T560MANK5 DOCX Figure 6 3 XPort Home Web Page Figure 6 4 XPort Network Settings ...
Страница 36: ...36 T560MANK5 DOCX Figure 6 5 XPort Server Settings Figure 6 6 XPort Serial Settings ...
Страница 42: ...42 T560MANK5 DOCX Figure 8 1 T560 Outline and Mounting ...
Страница 43: ...43 T560MANK5 DOCX Figure 8 2 Flange Mounting Dimensions ...
Страница 44: ...44 T560MANK5 DOCX Figure 8 3 Printed Circuit Board Dimensions ...