Perform a MAINS RESET:
− Switch off the mains switch (position "0", fig. 7.1, pos. 2).
− Wait at least 10 seconds and then switch on the mains switch again to position "1".
13.3
Changing the fuse
Switch off the mains switch and remove the power-plug from the external power-supply from the mains!
Srew the cap from the fuse-holder, fig. 7.1, pos. 5, off by turning it counter-clockwise, 1/8-turn, and pull it out with the
fuse. Exchange the defective fuse and tightly screw it including the cap back into the fuse-holder by turning it in
clockwise direction.
Only T10A/250V, 6.3 x 32 mm fuses with
R
- and
R
-approval
(order no. E2268) may be used.
14
Returning the device or some parts of it
If the device, some parts of it or its accessories are returned to Hettich AG or the local supplier, in order to
provide protection for people, the environment and materials, it has to be decontaminated and cleaned
before being shipped, a declaration should to be apply onto the device, respectively onto the parts.
Before returning the device, a transport securing part has to be installed.
Before returning the device or some parts of it, by some local suppliers you should ask for an RMA-
number (Return Authorization Number)
We reserve the right to refuse contaminated devices or accessories.
Costs incurred for cleaning and disinfection are to be charged to the customer.
We ask for your understanding in this matter.
15
Storage
Before storage the device, it should be decontaminated and cleaned to protect people, the environment
and property. We suggest to put a declaration onto the device with the date, sign and the solution.
The appliance may only be stored under the following conditions:
- in a dry and enclosed, dust-free room
- frost-free
- disconnected from the power supply
15.1
Disposal
Before disposal, the device must be decontaminated and cleaned to protect people, the
environment and property. Please also observe all other regulations applicable in this context.
We
suggest to apply a declaration onto the device with the date, sign and the solution.
When you are disposing of the device, the respective statutory rules must be observed.
Pursuant to guideline 2002/96/EC (WEEE), all devices supplied after August 13, 2005 may not be
disposed as part of domestic (household), or industrial waste. The device belongs to group 8 (medical
devices) and is categorized in the business-to-business field.
The icon of the crossed-out trash can shows that the device may not be disposed as part of domestic
waste.
The waste disposal guidelines of the individual EC countries might vary. If necessary, contact your
supplier.
There is a lithium battery in the main-electronic-board of the device. Remove it and dispose of it in
accordance with the regulations in your country, if necessary, contact your supplier.