12.6 

Glass breakage 

In a case of glass breakage, the pieces of glass and leaked centrifugal liquid must be carefully removed from 
the centrifugal area and tube holders. 

The leakage could involve infectious materials, the area must be disinfected immediately. 
The drain-hole from the bowl should be blocked, a.e. with a plug, pencil rubber or with a crimped over 
tissue, so that none piece of glass can fall down into the drain-hole and block the drain-outlet connector on 
the rear-side of the device, fig. 7.1, pos. 8, befor starting to remove all broken pieces of glass. 

 

Before continue to work  

 

The bowl must be checked for scratches, if so a technician must replace it => call your local 
distributor. 

 

The rotor and its tube-holders must be checked for scratches and the rotor must be checked for 
correct function, if so or you are unable to resolve that the rotor must be replaced. 

 

13 

Faults and Errors 

13.1 

Operator faults 

 

Fault 

Cause 

Remedy 

Incomplete wash cycle 

 

A 24-place rotor is being used, even 

  though a 12-place rotor has been 
  set as the rotor type. 

 

The filling volume is set too low. 

 
 

 

The injection nozzles are blocked. 

 

Check the 

ROTOR: 

parameter in 

  the system settings. 
 

 

Check the volume (ml) and speed 

  from the FILL-process in the used  
  program. 

 

Clean the injection nozzles. 

Pellet does not form on 
the base of used tube 

 

The revolutions per minute during 

  the centrifugation run for the 
  agglutination test are too low. 

 

The tube holders are left hanging in 

  the decantation position 

 

Check the spin-process from the used 

  program. 
 

 

Check the rotor functions. 

 

 

Wrong used tube-type tube 

 

Check if the used tube is the same as 

  during the program-validation 

No pellet/pellet too 
small 

 

A 12-place rotor is being used, even 

  though a 24-place rotor has been 
  set as the rotor type. 

 

The filling volume is set too high. 

 

Wrong used tube-type tube 

 

Check the 

ROTOR: 

parameter in 

  the program. 

 

Check the 

SALINE (ml) 

parameter 

  in the program. 

 

Check if the used tube is the same as 

  during the program-validation 

The liquid is not 
decanted. 

 

The rotor mechanism is defective. 

 

The revolutions per minute during 

  the DECANT-process are too low. 

 

Check the rotor functions 

 

Check the spin-process from the used 

  program.

 

 

13.2 

Software Error Codes 

 

Error 

Code 

Error Name 

Description 

Possible reasons 

No error 

Program was successful, no error has 
occurred 

 

Running 

Program is still running, so far no error has 
been detected (should never be visible in 
history) 

 

Motor error 

10 

Motor 
startup error 

Motor could not be started (no speed could be 
detected) 

 

Motor is blocked 

 

Motor cable connection problem 

 

Motor power supply problem 

11 

Motor 
acceleration 
error 

Motor could not accelerate within tolerance 
(motor was too slow) 

 

Wrong rotor type selected 

 

Mechanical friction too big 

Содержание 1008-00

Страница 1: ...2017 by Hettich AG page 1 CW II OM rev1 0 Waschzentrifuge Helmer UltraCW II Waschzentrifuge Hettich Rotolavit II Operating Instructions...

Страница 2: ...Centrifugeuse lave cellule Helmer UCW II 0000020 num ro de serie Dichiariamo sotto la nostra unica responsabilit che il dispositivo medico diagnostico in vitro Centrifuge per lavaggio Hettich Rotolav...

Страница 3: ...d valuation du risique Procedimento di valutazione del rischio EN ISO 14971 Konformit tsbewertungsverfahren Conformity assessment procedure Proc dure d valuation da la conformit Procedimento di valut...

Страница 4: ...AG Seestrasse 204a CH 8806 Baech Switzerland phone 41 0 44 786 80 20 info hettich ch www hettich ch 2017 by Hettich AG All rights reserved No part of this publication may be reproduced without the pri...

Страница 5: ...e of malfunctions and irregularities 11 4 1 Remaining risks 11 4 2 Switching off the device in an emergency 11 4 3 Emergency release 11 5 Technical specifications 12 6 Unpacking the centrifuge 13 6 1...

Страница 6: ...ss 25 9 4 7 AGIT process 25 9 4 8 LOOP process 26 9 4 9 CHECK process 26 9 5 Add a new program 27 10 Adjusting 29 10 1 Entering the rotor type 29 10 2 Calibrating the filling volume 29 10 3 Audible si...

Страница 7: ...2017 by Hettich AG page 7 CW II OM rev1 0 Hettich AG Laborapparate phone 41 44 786 80 20 Seestrasse 204a e mail info hettich ch CH 8806 Baech internet www hettich ch 16 1 Rotors and accessories 37...

Страница 8: ...al damage and injury to personal Symbol in this document This symbol refers to important circumstances Symbol on the device and in this document Beware of biohazard Symbol on the device and in this do...

Страница 9: ...less condition of the device and for its proper operation in accordance to the specifications is responsible for ensuring that persons who are to operate or service the device are qualified to do this...

Страница 10: ...and the legal regulations on accident prevention you should also follow the recognised professional regulations for working in a safe and professional manner These operating instructions should be re...

Страница 11: ...ions or damage immediately take the device out of work and inform your superior You can find on correcting malfunctions in chapter 13 4 1 Remaining risks The device is built according to the state of...

Страница 12: ...yes no yes optional car battery DC input no no 11V 30VDC width mm inch 330mm 13inch depth mm inch 480mm 18 9inch height mm closed 280mm 11inch height mm open 580mm 22 9inch weight kg lb 24 4 53 8 24...

Страница 13: ...nce for damage If you notice deviations from the delivery note damage or irregularities do not put the appliance into operation but inform the haulage company and your supplier If it is possible keep...

Страница 14: ...7 1 Connections 1 mounting bracket for the Power supply 2 ON OFF main switch 3 Ethernet interface 4 DC power input 1 A 5 fuse fuse holder 6 inlet 1 Saline 7 inlet 2 solution 2 1 8 outlet Drain A solut...

Страница 15: ...on fig 7 1 pos 6 and dip the other side with the metal part into the container with the saline solution If the fill hose is to short for your installation and it is not possible to move closer the ca...

Страница 16: ...creen occurs push the lid open button and open the lid chapter 8 2 pos 5 remove the transportation safety device from the top of the rotor and keep it on a save place 8 Operation settings 8 1 Overview...

Страница 17: ...process of the program x 2 edit process x 3 back without saving to select a program x 4 add process x 5 navigate through the existing process of the program x 6 rename the program x 8 5 System Setting...

Страница 18: ...em settings 2 last or next month the day with a yellow background has some saved datas runs 8 5 2 Time settings Description Login wheel adjuster for date and time 1 accept changes 2 discard changes an...

Страница 19: ...lid x 2 fill in solution refill undefined x 3 start calibration x 4 back to service menu without saving x 5 select pump x 6 select target for measurement x 7 indicate filling level x 8 target measure...

Страница 20: ...w hettich ch 8 6 3 Device settings Description Login 1 save changes x 2 back to service menu without saving x 3 change device name x 4 activate deactivate pumps x 5 adjust motor speed limit x 6 add ne...

Страница 21: ...utton Select a program To add a new program see chapter 0 Each program must be validated by the user with the custom tube If the tube will be changed then the program must be revalidated After select...

Страница 22: ...pin 30sec 3500 susp 3ml spin 30sec wash and anti 9 3 1 flush 1 This preinstalled system program is especially to flush the external and also the internal tubing system for that the first FILL process...

Страница 23: ...3500RPM DECANT 300RPM AGIT 19x LOOP 2x 9 3 4 agit and spin The process values are AGIT 19x SPIN 30sec 3500RPM 9 3 5 decant The process values are DECANT 300RPM 9 3 6 spin 30sec 3500 The process value...

Страница 24: ...ological saline solution is injected directly into the tubes in order to receive a good resuspension of the cells The best results are with a speed of 900RPM for both kind of rotor types The device wi...

Страница 25: ...number of revolutions per minute is adjustable from 0RPM to 2500RPM The excess is decanted at the selectable rotation speed During decanting the rotor turns in the opposite direction of the normal ro...

Страница 26: ...n add every other process except a LOOP Process that means you can program to wash out cells with two cycles unstead of tree with a DECANT process with a speed of a e 320RPM so that some small amount...

Страница 27: ...t the name push at the top In this example the first process is FILL 1 push to choose the FILL 1 symbol Below with the bar you can choose the filling capacity and the rotation speed at which it should...

Страница 28: ...add more process push By selecting a process you can switch the process to the previous process position To switch to left push Or after the next process position to right To edit a process select it...

Страница 29: ...underneath the injection tube inside from the lid and push the button start calibration cha 8 6 1 pos 3 Adjust the reading from the cylinder at the target measure cha 8 6 1 pos 8 with the buttons or...

Страница 30: ...rying operating hours Go into the system settings start screen cha 8 2 pos 2 Go into the service menu system settings cha 8 5 pos 2 11 Centrifugation of materials or mixtures of materials with a densi...

Страница 31: ...pter Removal of the drain gutter Check the filling volume of the saline solution see the Adjusting the filling volume chapter The system must be rinsed through with distilled water to avoid formation...

Страница 32: ...After removing the radioactive contaminants remove agent residue by rinsing with water only outside of the centrifuge or wipe off with a damp cloth The rotors and accessories must be dried directly a...

Страница 33: ...olume ml and speed from the FILL process in the used program Clean the injection nozzles Pellet does not form on the base of used tube The revolutions per minute during the centrifugation run for the...

Страница 34: ...cked Open or check button was pressed but the lid could not be opened Lid mechanically blocked 31 Unlocking failed Open or check button was pressed but the lock could not be unlocked Motor was still r...

Страница 35: ...incurred for cleaning and disinfection are to be charged to the customer We ask for your understanding in this matter 15 Storage Before storage the device it should be decontaminated and cleaned to p...

Страница 36: ...W II OM rev1 1 Hettich AG Laborapparate phone 41 44 786 80 20 Seestrasse 204a e mail info hettich ch CH 8806 Baech internet www hettich ch Log serialnumber ______________ inventorynumber _____________...

Страница 37: ...tube Kapazit t capacity ml 3 5 Ma e dimensions x L mm 10 x 75 12 x 75 Anzahl p Rotor number p rotor 12 12 Drehzahl speed RPM 3500 2 RZB RCF 1438 Radius radius mm 105 CW 1024 A E2197 Dekantierrotor 24...

Страница 38: ...ml 3 5 Ma e dimensions x L mm 10 x 75 12 x 75 Anzahl p Rotor number p rotor 12 12 Drehzahl speed RPM 3500 2 RZB RCF 1438 Radius radius mm 105 1018 A E2197 Dekantierrotor 24 fach Decant rotor 24 times...

Страница 39: ...Seestrasse 204a e mail info hettich ch CH 8806 Baech internet www hettich ch 16 1 3 Document History rev date supersession Revision description 1 0 23 May 2017 01 06 Contents desc of preloaded progra...

Отзывы: