![Heraeus Soluva Air W Скачать руководство пользователя страница 29](http://html.mh-extra.com/html/heraeus/soluva-air-w/soluva-air-w_instructions-manual_2125617029.webp)
IPS_Man_20003342641 V01 en 03.21
Page 29 of 36
Status
Description
Ready for operation
LED colour: green
Lighting mode: permanent
System is switched on, UV lamp preheating phase.
In operation
LED colour: blue
Lighting mode: permanent
System is switched on, UV lamps are active, no er-
ror.
System fails to start up
Main switch is set to ON, system does not react.
Root cause/action:
Connector not plugged in.
è
Check connector (plugged in?).
Fuse has tripped.
è
Open the system, switch the automatic circuit
breaker in the device on again (located at the
right next to the mains cable plug).
Customer's residual current circuit breaker in the
fuse cabinet has tripped due to an error.
è
Check customer's fuse cabinet.
UV lamp: Warning
LED colour: yellow
Lighting mode: permanent
System is switched on, UV lamps are active, the
end of the service life is near.
Root cause/action:
The end of the service life of the UV lamps is near.
è
Order lamp replacement kit.
UV lamp: Service life
LED colour: purple
Lighting mode: permanent
System is switched on, UV lamps are active, the
end of the service life is reached.
Root cause/action:
The service life of the UV lamps has expired.
è
Replace the UV lamps.
NOTICE! After lamp exchange, press the reset
button.
Reset
LED colour: white
Lighting mode: 5 seconds
Reset was carried out successfully after the button
has been kept pressed for five seconds.
Root cause/action:
Reset has been performed.