Disconnect the UV lamp plug connection (contacting).
è
Cautiously remove the UV lamp from the retainer.
Installing a new UV lamp
è
Establish the connection of the new UV lamp.
è
Insert the UV lamp into the retainer without canting it.
è
Tighten the knurled nuts of the lamp retainers.
è
Cautiously insert the lamp slide-in unit into the housing and screw it on.
Check the wiring when inserting the lamp slide-in unit.
If necessary, push projecting cables into the housing.
Do not route cables across the radiation area.
CAUTION! Make sure that cables are not squeezed by the hood.
Resetting the operating hours meter
è
Before closing the hood, switch the device on. This will not start the emission of UV
radiation.
Ensure that no unauthorised persons are within the immediate vicinity of the
device.
è
When a UV lamp is replaced, reset the operating hours meter with the
Reset operat-
ing hours
button.
To do this, keep the button pressed for about five seconds until the LED strip light is
lit up white.
è
Switch the device off before you close the hood.
è
Close the hood. The hood must engage audibly. The oblong holes of the hood must
completely surround the three screws at the device.
Correct
Wrong
è
Retighten all three screws.