3.1 Sterilization of detachable handpiece

3.1.1 Autoclaved to high temperature/pressure:

a) 

121°C/1bar0.1MPa.

b) 

135°C/2.2bar (0.22MPa).

c) 

Pull out the handpiece and unscrew the scaling tip and endochuck after each operation.

d) 

Pack the handpiece with sterile gauze or sterile bag before sterilizing.

e)

Reuse handpiece after it cools naturally in case of scalding hand.

3.1.2 Notice

a) 

Clean the cleaning liquid in the handpiece by compressed air before sterilization.

b) 

Be sure that the scaling tip has been unscrewed from the handpiece and it cannot be

sterilized with others.

c) 

Please notice whether the outer of the handpiece is damaged during the treatment or

sterilization. Don't smear any protective oil on the surface of handpiece.

d) 

There are two waterproof orings at the end of handpiece. Please lubricate them with

dental lube frequently, as sterilization and repeated pulling and inserting will reduce their
life-span. Change a new one once it is damaged or worn excessively.

e) 

The following sterilizing methods are forbidden:

Put handpiece into any liquid for boiling.
Dip handpiece in disinfectors such as iodine , alcohol or glutaraldehyde.
Put handpiece into oven or microwave oven for baking.

3.2 Sterilization of tips endo wrench, endochuck

All the scaling tips and endochuck can be autoclaved to 135°C.

3.3 Sterilization of torque wrench and endo wrench

a) 

The torque wrench and endo wrench can be sterilized under high temperature and

pressure.

b) 

The following sterilization ways for torque wrench are forbidden:

Braise in liquor.
Dip in iodine, alcohol or glutaraldehyde.
Torrefy in oven or microwave oven.

Notice:
We are not responsible for any damage of the torque wrench directly or
indirectly made by any way in the above items.

3.S

TERILISATION AND

M

AINTENANCE

V

ET

P

IEZO

S

CALER

PS1

3.4 Cleaning of tip, endochuck, torque wrench and endo wrench

Scaling tip, endochuck, torque wrench and endo wrench can be cleaned by ultrasonic cleaner.

3.5 Troubleshooting and notes

3.5.1 Troubleshooting

If the problem still can't be solved, please contact with local dealer or manufacturer.

3.S

TERILISATION AND

M

AINTENANCE

V

ET

P

IEZO

S

CALER

PS1

FAULT

POSSIBLE CAUSE

SOLUTIONS

The power pipe plug is in loose contact

The foot  switch is in loose contact

Insert the foot switch to its socket tightly

Contact our dealers or us

Contact our dealers or us

Screw  the tip on the handpiece tightly

(1.4.1 f))

Contact our dealers or us

Send it to our company to repair

Contact our dealers or us

Turn on the water control knob note 1

Contact our dealers or us

Clean the water pipe by

multi-function syringe note 2

Contact our dealers or us

Turn the water control knob

to a higher grade note 2

Make the water pressure higher

Clean the water pipe by

multi-function syringe note 2

Screw  the tip on the handpiece 

tightly (as showed in 1.4.1 f))

Screw on the tip tightly

(as showed in 1.4.1 f))

Dry it by the use of hot air

Change by a new one

Change it by a new "O" ring

Screw it tighly

Change it by a new one

Screw it tightly

Make the plug insert to the socket well

The fuse of transformer is  broken

The fuse in the main unit  is broken

The tip is in loose contact

The connect plug between the handpiece

and the circuit board is in loose contact

Something wrong with the handpiece

Something wrong with the cable

The water control knob is off

There is impurity in the 

electric-magnetic valve

The water pipe is blocked

There is impurity in the 

electric-magnetic valve

The water control knob is in 

a low setting

The water pressure is not high enough

The water pipe is blocked

The tip hasn't been screwed 

on to the handpiece tightly

The tip is shaken loose by vibration

The coupling between the handpiece

and the cable is wet

The tip is damaged note 3

The waterproof "O" ring is damaged

The screw cap hasn't been screw tightly

Endochuck is damaged

The screw cap hasn't been screw tightly

The scaling tip doesn't

vibrate and there’s no water

flowing out when stepping

on the foot switch

The scaling tip doesn't

vibrate but there is water

flowing out when stepping

on the foot switch.

The scaling tip vibrates but

there is no spray when 

stepping on the foot switch.

There is still water flowing

out after the power is off.

The  handpiece

The amount of spouting

water is too little

The vibration of the tip

becomes weak

Water seeping from the coupling

between handpiece and cable

The u-file doesn't vibrate

Noise coming 

from the endochuck

E

N

G

L

IS

H

Содержание PS1

Страница 1: ...hanical including photocopying typing or information retrieval systems without written permission of Henry Schein Inc Todos los derechos reservados Ninguna parte de estas instrucciones puede ser reproducida o copiada de ninguna manera y por ningún medio gráfico electrónico o mecánico incluyendo el fotocopiado tipeo o los sistemas de recuperación de informa ción sin la autorización expresa por escr...

Страница 2: ...ct is mainly used for teeth cleaning and is also an indis pensable equipment for tooth disease prevention and treatment The new product Vet Piezo Scaler PS1 has scaling endo automatic water supply function It contains the following features 1 Automatic frequency tracking ensures that the machine always works on the best fre quency and performs more steadily The optional function automatic water su...

Страница 3: ... 250VT 1 6AL h Power source fuse 250VT 0 5AL i Water pressure 0 1bar to 5bar 0 01MPa to 0 5MPa j Weight of main unit 0 73kg k Weight of power source 1 2kg l Operating mode Continuous operation m Type of protection against electric shock Class II n Degree of protection against electric shock Type B equipment o Degree of protection against harmful ingress of water Ordinary equipment IPX0 p Protectio...

Страница 4: ...t Foot switch socket Plug of foot switch Power supply Input voltage 220V to 230V Power cord plug Socket for power supply Power cord plug Connect pure water Cap of connector Water pipe Water entrance Switch for automatic water supply optional Air exit optional Connect water entrance small Connect air exit big ð ð ð ð Water entrance Reservoir Torque wrench 1 THEINSTALLATIONANDCOMPONENTSOFEQUIPMENT V...

Страница 5: ...TION VET PIEZO SCALER PS1 c HandpieceThe main part of the whole handpiece can be autoclaved under the high temperature and pressure d The connector of the cable Connect the handpiece with the water source and power supply of the main unit Notice Keep the joint of handpiece and the cable connector dry 2 1 3 Instruction of using torque wrench showed in 1 4 1 f a The torque wrench s structure is desi...

Страница 6: ...TENANCE VET PIEZO SCALER PS1 FAULT POSSIBLE CAUSE SOLUTIONS The power pipe plug is in loose contact The foot switch is in loose contact Insert the foot switch to its socket tightly Contact our dealers or us Contact our dealers or us Screw the tip on the handpiece tightly 1 4 1 f Contact our dealers or us Send it to our company to repair Contact our dealers or us Turn on the water control knob note...

Страница 7: ... on the foot switch 4 1 4 The scaling tip must be fastened and there must be fine spray or drip coming out form the tip when operating 4 1 5 Change by a new one when the tip and ultrasonic file are damaged or worn excessively 4 1 6 Don t twist or rub the tip and endochuck 4 1 7 Don t use impure water source and be sure not use normal brine instead of pure water source 4 1 8 If use the water source...

Страница 8: ...tation 4 4 3 Avoid solarization and getting wet in rain or snow during transportation We offer one year s free repair to the equipment according to the warranty card The repair of the equipment should be carried out by Henry Schein or an accred ited repair centre technician by Henry Schein We are not responsible for any irretrievable damage caused by a non accredited repair centre technician 4 PRE...

Страница 9: ...len Endo automatisches Wasserversorgungs System optional und LED Funktionen optional Es beinhaltet die folgenden Eigenschaften 1 Das automatische Verfolgen von Frequenzen stellt sicher dass das Gerät immer im besten Frequenzbereich und gleichmäßig arbeitet Optionale Funktion Automatisches Wasserversorgungs System MAGENTA OUT 2 Das Handstück ist abnehmbar und kann bis zu einer Maximaltemperatur von...

Страница 10: ...sgangsleistung 3W bis 20W g Sicherung Hauptgerät 250VT 1 6AL h Sicherung Leistungsquelle 250VT 0 5AL i Wasserdruck 0 1 bar bis 5 bar 0 01MPa bis 0 5MPa j Gewicht Hauptgerät 0 73kg k Gewicht Leistungsquelle 1 2kg l Betriebsart Dauerbetrieb m Art des Schutzes gegen Elektroschock Klasse II n Grad des Schutzes gegen Elektroschock Typ B Ausrüstung o Grad des Schutzes gegen gesundheitsgefährdenden Zuflu...

Страница 11: ...hse Fußschalter Stecker Fußschalter Netzteil Eingangsspannung 220V bis 230V Stecker Stromkabel Buchse Stromanschluss Stecker Stromkabel Destilliertes Wasser onnect pure water Anschlusskappe anschließen Wasserschlauch Wassereingang Schalter für automatische Wasserversorgung optional Luftausgang optional Wassereingang klein anschließen Luftausgang groß anschließen ð ð ð ð Wassereingang Reservoir Dre...

Страница 12: ...nter hoher Temperatur und Druck autoklaviert werden d Kabelsteckverbindung Schließen Sie das Handstück an die Wasserversorgung und Energieversorgung des Hauptgeräts an Hinweis Halten Sie das Anschlussstück des Handstücks und die Kabelsteck verbindung trocken 2 1 3 Anleitung zur Benutzung des Drehmomentschlüssels gezeigt in 1 4 1 f a Die Struktur des Drehmomentschlüssels ist so angelegt dass die St...

Страница 13: ...werden kann nehmen Sie bitte Kontakt zu Ihrem Händler oder Hersteller auf 3 STERILISIEREN UND WARTUNG VET PIEZO SCALER PS1 FEHLER MÖGLICHE URSACHE LÖSUNGEN Der Netzstecker steckt nicht richtig in der Steckdose Der Fußschalter ist nicht richtig eingesteckt Stecken Sie den Fußschalter richtig in die Dose Kontaktieren Sie uns oder unsere Händler Kontaktieren Sie uns oder unsere Händler Schrauben Sie ...

Страница 14: ...ußschalter betätigen 4 1 4 Die Scale Spitze muss fest sitzen und ein feiner Sprühstrahl oder Tröpfchen müssen während der Bedienung aus der Spitze kommen 4 1 5 Bitte wechseln Sie die Spitze und Ultraschall Feile aus wenn diese beschädigt oder extrem abgenutzt sind 4 1 6 Verbiegen oder polieren Sie die Spitze und das Endo Bohrfutter nicht 4 1 7 Benutzen Sie kein unreines Wasser oder gewöhnliche Lau...

Страница 15: ...ansportieren Sie das Gerät nicht zusammen mit Gefahrgütern 4 4 3 Vermeiden Sie während des Transports Sonneneinstrahlung und Feuchtigkeit durch Regen oder Schnee Gemäß der Garantiekarte bieten wir ein Jahr gratis Reparaturen für dieses Gerät an Reparaturen am Gerät sollten durch unseren Fachtechniker durchge führt werden Wir sind nicht verantwortlich für irreversible Schäden die durch Nicht Fachle...

Отзывы: