2.1 Scaling function.

2.1.1 Operation.

a) 

Open the packing box, make sure that all the parts and accessories are complete

according to the packing list. Take the main unit out of the box and put it on a stable plane.

b) 

Turn the water control knob to the max based on symbol as shown as 3.5.2 [note 1].

c) 

Insert the plug of the foot switch to its socket (1.4.1 b).

d) 

Connect one end of the water pipe to the water entrance, and the other end to the clean

water source (1.4.1 c).

e) 

Screw the scaling tip tightly to handpiece by torque wrench , then connect the hand-

piece and the connector of cable correctly.

f) 

Connect the main unit with output plug of power source, then connect to the  power

(1.4.1 b).

g) 

Switch on the main unit, then the scaling indicator and the first five lead lights of

power indicator shine.

h) 

Select a suitable scaling tip as you need,screw it on the handpiece tightly by the

torque wrench (1.4.1 f).

i) 

The normal frequency is extremely high. Under the normal working state of the scal-

ing tip, a light touch and a certain to-and-fro motion will eliminate the tartar without
heating. Overexertion and long-time lingering are forbidden.

j) 

Vibrating intensity: Adjust the vibration intensity as you need , generally turn the

knob to the middle grade. According to patient's different sensitivity and the rigidity of
the gingival tartar,adjust the vibration intensity during the clinical treatment.

k) 

Water volume adjustment: Step on the foot switch, and the tip begins to vibrate, then

turn the water control knob to form fine spray to cool down the handpiece and clean
the teeth.

l) 

The handpiece can be handled in the same gesture as a pen in hand.

m) 

During the clinical treatment, be sure not to make the end of tip touch the teeth ver-

tically and not to make the tip overexert on the surface of the teeth in case of hurting
the teeth and damaging the tip.

n) 

After finishing operation, keep the machine working for 30 seconds on the water

supply condition, to clean the handpiece and the scaling tip.

o) 

Unscrew the scaling tip and pull out handpiece, then sterilize them.

Notice: Don't pull out the handpiece when the foot switch is stepped on and
the machine is working.

2.1.2 Instruction for main components of detachable handpiece
(showed in 1.4.1 e).

a) 

NippleThe nipple can be removed.You can screw off the nipple and clean the pole with

alcohol termly.

b) 

Handpiece decorative ring. The seal can be removed and cleaned with alcohol termly.

2.P

RODUCT FUNCTION AND OPERATION

V

ET

P

IEZO

S

CALER

PS1

c) 

HandpieceThe main part of the whole handpiece, can be autoclaved under the high

temperature and pressure.

d) 

The connector of the cable Connect the handpiece with the water source and power

supply of the main unit.

Notice: Keep the joint of handpiece and the cable connector dry.

2.1.3 Instruction of using torque wrench (showed in 1.4.1 f).

a) 

The torque wrench's structure is designed in special way which can control the strength

of the scaling tip installation properly and correctly. It also can guarantee the operator screw
or unscrew the scaling tip effectively and keep their hands away from being scratched.

b) 

Operation: Take the scaling tip into the torque wrench, operate as showed in 1.4.1 f).

Tip installation: Hold the handpiece, turn the tip toward direction as showed in 1.4.1 f) with
the torque wrench. Turn one more circles when the tip stops, then the tip is installed.
Tip uninstallation: Hold the handpiece, turn the wrench toward anti-clockwise direction.
Sterilize it in sterilizer after each treatment. The torque wrench must be cooled naturally
after sterilization to avoid scalding when using next time. Keep the torque wrench in a cool,
dry and ventilated place and keep it clean.

2.1.4 Automatic water supply system(optional).

a) 

Usage: After adding enough water into the reservoir, install the reservoir cap on the

reservoir properly and tighten it up.
The twin tube is composed of two pipes. The big one is air pipe, and small one is water
pipe. Connect the air pipe to AIR connector on the reservoir, connect the water pipe to
H2O  connector. Turn on the automatic water supply switch on the main unit.

b) 

Precaution: Please operate correctly according to  the manual, the reservoir cap must

be tighten up. When adding or changing water, please pull out the air pipe first, then pull
out the water pipe. Under normal automatic water supply condition,the air pump produces
the WOO sound intermittently, which is common phenomenon.

2.2 Endo function

a) Usage process:

Fix endochuck to handpiece by endo wrench.(See 1.4.1 f)). Unscrew the screw cap on
the endochuck. Put the ultrasonic file into the hole in the front of endochuck. Screw down
the screw cap with endo wrench, to tight up the ultrasonic file. Press option  key, turn to
endo function, then the indicator of endo function is on. When ultrasonic scaler turns into
endo function,only the first lead light is on and the power is at the first grade. Put the ultra-
sonic file into the patient's root canal slowly, step on the foot switch, then make endo treat-
ment. During the treatment, turn up the power gradually according to the needs.

2.P

RODUCT FUNCTION AND OPERATION

V

ET

P

IEZO

S

CALER

PS1

E

N

G

L

IS

H

Содержание PS1

Страница 1: ...hanical including photocopying typing or information retrieval systems without written permission of Henry Schein Inc Todos los derechos reservados Ninguna parte de estas instrucciones puede ser reproducida o copiada de ninguna manera y por ningún medio gráfico electrónico o mecánico incluyendo el fotocopiado tipeo o los sistemas de recuperación de informa ción sin la autorización expresa por escr...

Страница 2: ...ct is mainly used for teeth cleaning and is also an indis pensable equipment for tooth disease prevention and treatment The new product Vet Piezo Scaler PS1 has scaling endo automatic water supply function It contains the following features 1 Automatic frequency tracking ensures that the machine always works on the best fre quency and performs more steadily The optional function automatic water su...

Страница 3: ... 250VT 1 6AL h Power source fuse 250VT 0 5AL i Water pressure 0 1bar to 5bar 0 01MPa to 0 5MPa j Weight of main unit 0 73kg k Weight of power source 1 2kg l Operating mode Continuous operation m Type of protection against electric shock Class II n Degree of protection against electric shock Type B equipment o Degree of protection against harmful ingress of water Ordinary equipment IPX0 p Protectio...

Страница 4: ...t Foot switch socket Plug of foot switch Power supply Input voltage 220V to 230V Power cord plug Socket for power supply Power cord plug Connect pure water Cap of connector Water pipe Water entrance Switch for automatic water supply optional Air exit optional Connect water entrance small Connect air exit big ð ð ð ð Water entrance Reservoir Torque wrench 1 THEINSTALLATIONANDCOMPONENTSOFEQUIPMENT V...

Страница 5: ...TION VET PIEZO SCALER PS1 c HandpieceThe main part of the whole handpiece can be autoclaved under the high temperature and pressure d The connector of the cable Connect the handpiece with the water source and power supply of the main unit Notice Keep the joint of handpiece and the cable connector dry 2 1 3 Instruction of using torque wrench showed in 1 4 1 f a The torque wrench s structure is desi...

Страница 6: ...TENANCE VET PIEZO SCALER PS1 FAULT POSSIBLE CAUSE SOLUTIONS The power pipe plug is in loose contact The foot switch is in loose contact Insert the foot switch to its socket tightly Contact our dealers or us Contact our dealers or us Screw the tip on the handpiece tightly 1 4 1 f Contact our dealers or us Send it to our company to repair Contact our dealers or us Turn on the water control knob note...

Страница 7: ... on the foot switch 4 1 4 The scaling tip must be fastened and there must be fine spray or drip coming out form the tip when operating 4 1 5 Change by a new one when the tip and ultrasonic file are damaged or worn excessively 4 1 6 Don t twist or rub the tip and endochuck 4 1 7 Don t use impure water source and be sure not use normal brine instead of pure water source 4 1 8 If use the water source...

Страница 8: ...tation 4 4 3 Avoid solarization and getting wet in rain or snow during transportation We offer one year s free repair to the equipment according to the warranty card The repair of the equipment should be carried out by Henry Schein or an accred ited repair centre technician by Henry Schein We are not responsible for any irretrievable damage caused by a non accredited repair centre technician 4 PRE...

Страница 9: ...len Endo automatisches Wasserversorgungs System optional und LED Funktionen optional Es beinhaltet die folgenden Eigenschaften 1 Das automatische Verfolgen von Frequenzen stellt sicher dass das Gerät immer im besten Frequenzbereich und gleichmäßig arbeitet Optionale Funktion Automatisches Wasserversorgungs System MAGENTA OUT 2 Das Handstück ist abnehmbar und kann bis zu einer Maximaltemperatur von...

Страница 10: ...sgangsleistung 3W bis 20W g Sicherung Hauptgerät 250VT 1 6AL h Sicherung Leistungsquelle 250VT 0 5AL i Wasserdruck 0 1 bar bis 5 bar 0 01MPa bis 0 5MPa j Gewicht Hauptgerät 0 73kg k Gewicht Leistungsquelle 1 2kg l Betriebsart Dauerbetrieb m Art des Schutzes gegen Elektroschock Klasse II n Grad des Schutzes gegen Elektroschock Typ B Ausrüstung o Grad des Schutzes gegen gesundheitsgefährdenden Zuflu...

Страница 11: ...hse Fußschalter Stecker Fußschalter Netzteil Eingangsspannung 220V bis 230V Stecker Stromkabel Buchse Stromanschluss Stecker Stromkabel Destilliertes Wasser onnect pure water Anschlusskappe anschließen Wasserschlauch Wassereingang Schalter für automatische Wasserversorgung optional Luftausgang optional Wassereingang klein anschließen Luftausgang groß anschließen ð ð ð ð Wassereingang Reservoir Dre...

Страница 12: ...nter hoher Temperatur und Druck autoklaviert werden d Kabelsteckverbindung Schließen Sie das Handstück an die Wasserversorgung und Energieversorgung des Hauptgeräts an Hinweis Halten Sie das Anschlussstück des Handstücks und die Kabelsteck verbindung trocken 2 1 3 Anleitung zur Benutzung des Drehmomentschlüssels gezeigt in 1 4 1 f a Die Struktur des Drehmomentschlüssels ist so angelegt dass die St...

Страница 13: ...werden kann nehmen Sie bitte Kontakt zu Ihrem Händler oder Hersteller auf 3 STERILISIEREN UND WARTUNG VET PIEZO SCALER PS1 FEHLER MÖGLICHE URSACHE LÖSUNGEN Der Netzstecker steckt nicht richtig in der Steckdose Der Fußschalter ist nicht richtig eingesteckt Stecken Sie den Fußschalter richtig in die Dose Kontaktieren Sie uns oder unsere Händler Kontaktieren Sie uns oder unsere Händler Schrauben Sie ...

Страница 14: ...ußschalter betätigen 4 1 4 Die Scale Spitze muss fest sitzen und ein feiner Sprühstrahl oder Tröpfchen müssen während der Bedienung aus der Spitze kommen 4 1 5 Bitte wechseln Sie die Spitze und Ultraschall Feile aus wenn diese beschädigt oder extrem abgenutzt sind 4 1 6 Verbiegen oder polieren Sie die Spitze und das Endo Bohrfutter nicht 4 1 7 Benutzen Sie kein unreines Wasser oder gewöhnliche Lau...

Страница 15: ...ansportieren Sie das Gerät nicht zusammen mit Gefahrgütern 4 4 3 Vermeiden Sie während des Transports Sonneneinstrahlung und Feuchtigkeit durch Regen oder Schnee Gemäß der Garantiekarte bieten wir ein Jahr gratis Reparaturen für dieses Gerät an Reparaturen am Gerät sollten durch unseren Fachtechniker durchge führt werden Wir sind nicht verantwortlich für irreversible Schäden die durch Nicht Fachle...

Отзывы: