Prohibit drop from high space. 

Under the high temperature or pressure is not allowed to use, if high 
humidity, then need to add the device to decrese the humidity. 

Human damage is not free repairment. 

Befor  enter  into  dangerous  area,  people  need  to  be  electrostatic 

discharged, then bring the detector    into the working place.   

Using the special charger for this detector. 

8.Common faults and treatments

 

 

  Value display not stable, up and down 

      Reason

some  electrochemical  sensor  easily  effected  by  other  gas, 

maybe is colorless and odorless. 

      Solution

Bring the detector into clean air, then the data will decrease to 

normal value; If still not normal, then zero calibrate in clean air. 

 

Response weak 

      Reason

Lower level of oxygen

<5%VOL

 

High pressure, pump not work

 

                          The sensor useful life end, or other faults

 

      Solution

If  the  Electrochemical  sensor,  Catalytic  combustion 

sensor,Semiconductor  sensor,need  oxygen  to  work,  so the 
level of oxygen should more than 5%VOL

pressure range 

is

-30Kpa 

  100Kpa

If  user  have  calibrate  gas,  can 

import the gas to test and target point calibrate. If oxygen 
and pressure both normal, then maybe other faults, need to 
send back to manufacturer.                 

 

 

After access the detecting gas, the data cann’t stable for a long time, up 

and down 
      Reason

Maybe the too low content of oxygen in the detecting gas, or 

the change of the detecting gas. 

      Solution

Increasing  the  oxygen  level  and  keep  stable  flow  rate,  or 

Содержание OC-1000

Страница 1: ...OC 1000 Gas and dust particle detector Operating Manual Henan Oceanus Import Export Co Ltd...

Страница 2: ...1 Catalogue Notice 2 Description 4 Feature 4 Parameter 5 Structure 7 Operation instruction 7 Operation interface 9 Remarks 12 Fault Treatment 13 Data logger funct ion 14...

Страница 3: ...ss 5s on off Press 5s Print Print real time data Attentions OC 1000 portable gas detector only avaible to detect the gas concentration when the pump on Treatment for the over measuring range Please av...

Страница 4: ...g the free calibration service once in a year About the Backlight instruction OC 1000 with the electricity saver procedure when the device without any operation within 30s the detector will turn on th...

Страница 5: ...PS industrial grade color screen displaying each parameter It is able to select gas concentration unit of PPM VOL mg m3 For 1 5 gas detecting supporting PM2 5 dust particle sensor and temperature sens...

Страница 6: ...Gas type At most for 5 sensors the gas type will be according to your requirements Temperature and humidity Temperature 40 120 Humidity 0 100 RH Testing mode Pump sunction inner pump adjustable flow...

Страница 7: ...rget Gas List Some gas not list if need pls contact manufacturer Gas Range Accuracy Resolution Reponse time Ex 0 100 LEL 3 F S 0 1 LEL 20s CH4 0 100 LEL 3 F S 0 1 LEL 20s O2 0 30 VOL 3 F S 0 01 VOL 20...

Страница 8: ...7 4 Structure 5 Operation instruction 5 1 Keys Total seven keys Back OK run stop and print Three interfaces Testing interface Menu interface Setting interface...

Страница 9: ...off Press 5s on off Press 5s Print Print real time data Attentions OC 1000 portable gas detector only avaible to detect the gas concentration when the pump on 5 2 Power on When power off long press ke...

Страница 10: ...nt interface as the following picture displaying one gas two gas three gas four gas orderly As the above displaying time temperature pump running battery and humidity Take A for example for O3 CO CH2O...

Страница 11: ...an turn on or off some one or more channel gas With the Pump flow rate can adjust the pump flow rate only need to setting under vaccum enviroment default is 30 With Temperature correction and Humidity...

Страница 12: ...re enter into the target gas calibration interface slowly turn on the calibrate gas and control the flow rate within 500ml min observing the detector gas concentration now should be rising when the va...

Страница 13: ...n is this period can set the timing cycle and round cycle time between the twice timing detecting Setting the timing detection need to back to the Detecting interface to press the Run stop key turn on...

Страница 14: ...se to normal value If still not normal then zero calibrate in clean air Response weak Reason Lower level of oxygen 5 VOL High pressure pump not work The sensor useful life end or other faults Solution...

Страница 15: ...if not then need to change the charger if is 5V then need to send back to manufacturer for testing 9 Data logger OC 1000 with the data logger function the charging port is the data output port the ope...

Страница 16: ...be in this period range 6 Click the search can find the above period gas concentration for the according channel gas the data also could be saved as the EXCEL form by click the Read Data Remark Data...

Отзывы: