background image

14 

 

change the higher oxygen calibrate gas. 

 

Pump weak, lack strength 

      Reason

Pump  have  suction  many  particle  and  gas,  the  entrance 

blocked. 
      Solution

Send back to manufacturer to change a new pump, adding the 

dust filter in the later usage.

 

 

Turn on fails 

      Reason

No power. 

Solution

After  charge,  if  still  cann’t  turn  on  it,  then  send  back  to 

manufacturer. 

 Charged fails 

      Reason

Charger 

problems 

or 

wrong 

model 

charger

Usually5-5.5VDC

1-2A

 

Solution

Check the charger output voltage if is 5V,if not, then need to 

change  the  charger,  if  is  5V,  then  need  to  send  back  to 
manufacturer for testing. 

9.Data logger 

        OC-1000  with  the  data  logger  function,  the  charging  port  is  the 

data output port, the operation steps as the following: 

        1

Install the export software on the computer. 

        2

Turn on the detector, after the 60s preheating over, connectted 

with the computer.   

        3

After install the drive of communication line, open the software, 

and confirm the computer COM port, then click the “Open 
port”,  now  the  lower  left  corner  will  display  the  “Port 
Open ”(Baud rate default is 115200) 

        4

Enter  into  the  history  data  menu  interface,  export  the  gas 

concentration data, if complete will appear the cues.   

               

Содержание OC-1000

Страница 1: ...OC 1000 Gas and dust particle detector Operating Manual Henan Oceanus Import Export Co Ltd...

Страница 2: ...1 Catalogue Notice 2 Description 4 Feature 4 Parameter 5 Structure 7 Operation instruction 7 Operation interface 9 Remarks 12 Fault Treatment 13 Data logger funct ion 14...

Страница 3: ...ss 5s on off Press 5s Print Print real time data Attentions OC 1000 portable gas detector only avaible to detect the gas concentration when the pump on Treatment for the over measuring range Please av...

Страница 4: ...g the free calibration service once in a year About the Backlight instruction OC 1000 with the electricity saver procedure when the device without any operation within 30s the detector will turn on th...

Страница 5: ...PS industrial grade color screen displaying each parameter It is able to select gas concentration unit of PPM VOL mg m3 For 1 5 gas detecting supporting PM2 5 dust particle sensor and temperature sens...

Страница 6: ...Gas type At most for 5 sensors the gas type will be according to your requirements Temperature and humidity Temperature 40 120 Humidity 0 100 RH Testing mode Pump sunction inner pump adjustable flow...

Страница 7: ...rget Gas List Some gas not list if need pls contact manufacturer Gas Range Accuracy Resolution Reponse time Ex 0 100 LEL 3 F S 0 1 LEL 20s CH4 0 100 LEL 3 F S 0 1 LEL 20s O2 0 30 VOL 3 F S 0 01 VOL 20...

Страница 8: ...7 4 Structure 5 Operation instruction 5 1 Keys Total seven keys Back OK run stop and print Three interfaces Testing interface Menu interface Setting interface...

Страница 9: ...off Press 5s on off Press 5s Print Print real time data Attentions OC 1000 portable gas detector only avaible to detect the gas concentration when the pump on 5 2 Power on When power off long press ke...

Страница 10: ...nt interface as the following picture displaying one gas two gas three gas four gas orderly As the above displaying time temperature pump running battery and humidity Take A for example for O3 CO CH2O...

Страница 11: ...an turn on or off some one or more channel gas With the Pump flow rate can adjust the pump flow rate only need to setting under vaccum enviroment default is 30 With Temperature correction and Humidity...

Страница 12: ...re enter into the target gas calibration interface slowly turn on the calibrate gas and control the flow rate within 500ml min observing the detector gas concentration now should be rising when the va...

Страница 13: ...n is this period can set the timing cycle and round cycle time between the twice timing detecting Setting the timing detection need to back to the Detecting interface to press the Run stop key turn on...

Страница 14: ...se to normal value If still not normal then zero calibrate in clean air Response weak Reason Lower level of oxygen 5 VOL High pressure pump not work The sensor useful life end or other faults Solution...

Страница 15: ...if not then need to change the charger if is 5V then need to send back to manufacturer for testing 9 Data logger OC 1000 with the data logger function the charging port is the data output port the ope...

Страница 16: ...be in this period range 6 Click the search can find the above period gas concentration for the according channel gas the data also could be saved as the EXCEL form by click the Read Data Remark Data...

Отзывы: