53
Immunity test
IEC 60601 test level
Compliance level Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
See NOTE 2 and NOTE 3
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the HemoCue systems, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter
Recommended separation distance
d=1.2√P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
Where (P) is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and (d) is the recom-
mended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey (a),
should be less than the compliance level in each
frequency range (b).
Interference may occur in the vicinity of equip-
ment marked with the following symbol.
Содержание Glucose 201+
Страница 1: ...HemoCue Glucose 201 Operating Manual Bedienungsanleitung Manuel de l utilisateur Gebruiksaanwijzing ...
Страница 2: ......
Страница 6: ...6 Components Bestandteile Composants Onderdelen ...
Страница 8: ...8 1a 3 1c 5 1b 2 4 6 Start up Einschalten Démarrage Opstarten ...
Страница 12: ...12 Measuring Capillary blood Messung Kapillarblut Mesure Sang capillaire Meting Capillair bloed 1 2 3 4 ...
Страница 14: ...14 5 6 7 8 ...
Страница 16: ...16 9 10 11 ...
Страница 18: ...18 15 16 13 12 15 14 ...
Страница 22: ...22 4 1 2 5 6 3 Set up QC test Einstellung QC Test Configurer Test QC Set up QC test ...
Страница 28: ...28 2 3 1 Set up Printer function Einstellung Druckerfunktion Configurer Mode impression Set up Printerfunctie ...
Страница 82: ......
Страница 83: ......