Quick Installation Guide
Document
WERA4-Quick-US-200917
Page 46 of 73
7.1.1
Acquisition Modes
STOP
Stop any running cycles after they finish; deactivate automatic
cycle control; Interrupt/Stop full calibration cycles (if it takes too
long); stop test modes; required before other modes can be
selected
Continuous Acquisition
Start continuous acquisition cycles as defined in WeraDesk
Permanent Acquisition
Start permanent acquisition cycles as defined in WeraDesk
Infinite Acquisition
Start infinite acquisition cycles as defined in WeraDesk
Single Acquisition
Start one acquisition cycle immediately
Calibration
Start one calibration cycle immediately
Cont. Calibration
Start continuous calibration cycles as defined in WeraDesk
Internal Calibration
Start an internal calibration cycle
Full Calibration
Start a full calibration cycles - caution that takes ~ 5 hours
Internal Full Calibration
Start internal full calibration cycles - caution that takes ~ 5 hours
FM RAW Acquisition
Run single measurement in chirp mode without internal range
resolution. RAW data will be saved.
Cont. FM RAW Acquisition
Start continuous FM raw acquisition cycles
Test Acquisition
Switch WERA to chirped mode without storing data; can
immediately be stopped
Test Calibration
Switch WERA to non-chirped (single tone) mode without storing
data; can immediately be stopped
Test internal Calibration
Switch WERA to internal calibration mode without storing data; can
immediately be stopped
Scan Frequencies
Activate a single frequency scan as defined in the frequency
management fields
Remarks:
-
To dynamically update the 'Maximum Range' and 'Data Acquisition Time' fields, Javascript needs to
be enabled for the browser. Do not enter values into these two fields.
-
If you allow JavaScript to open unrequested windows, a window showing the current WERA status
will automatically pop up. (Allow pop-ups for your browser)
-
Clicking 'Open Status Window' opens the status window manually. For updating the frequency scan
image of the status page the cache of some browsers (e.g. Mozilla and Firefox) must be disabled by
setting the cache size to zero. If the frequency scan image is not updated by the browser, manually
update the web page e.g. by pressing the F5 key under Firefox or Internet Explorer.
7.1.2
Interrupt Acquisition
By selecting this check-
box and submitting a “stop” as acquisition mode, any type of
acquisition will immediately interrupted. Automatic cycles will be deactivated.
Note that any full calibration measurement is a series of 256 single calibration
measurements. By using the interrupt Acquisition check box only the currently running
calibration measurement will be stopped.