Helmer Scientific i.Series

®

 Ultra-Low Freezer

Operation Manual 

360172-A/D

16

Product Specifications

6.1 

Operating Standards

 

Indoor use only

 

♦ Altitude (maximum): 2000 m

 

♦ Ambient temperature range: 15 °C to 32 °C

 

♦ Relative humidity (maximum for ambient temperature): 80% for temperatures up to 25 °C, decreasing linearly to 53% at 32 °C

 

♦ Temperature control range: -50 °C to -86 °C

 Note

Power conditioning (voltage boost) is an optional feature. If facility voltage is consistently at or below 195 V, voltage boost must 

be installed in order to protect the compressors.

Table 2.  Electrical Specifications

Model

116

118

124

126

Input Voltage and Frequency 

208/230 V, 60 Hz

Voltage Tolerance

±10%

Circuit Breakers

12.0 A (quantity 2)

Current Draw

11.0 A

Energy Consumption

18.5 kWh/day

18.5 kWh/day

19 kWh/day

19 kWh/day

Power Source

15 A dedicated circuit

Boost Cut-in Voltage

195 V

Monitoring System Battery /  
Chart Recorder Battery

12 V, 7 Ah rechargeable sealed lead acid battery

Remote Alarm Capacity

0.5 A at 30 V (RMS); 1.0 A at 60 V (DC)

 CAUTION

•  The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm system(s)  
  that uses normally-open or normally-closed dry contacts.
• 

If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s

 

  circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.

 Note

Overall exterior dimensions include casters, handle, i.C

3

 bezel, and door hinges

Table 3.  Ultra-Low Freezer Specifications

Model

Voltage Code

Amps

Capacity

 

Cu. Ft 

(Liters)

Dimensions Interior 

W x H x D in. (mm)

Dimensions 

W x H x D in. (mm)

Net Wt. 

lbs (kg)

Exterior

Overall Exterior

iUF116

208/230 V 

 

60 Hz

11.0

16

(453)

23.1 x 49.5 x 23.3 

(587 x 1257 x 592)

33.8 x 78.2 x 34.8 

(859 x 1986 x 884)

37.1 x 78.2 x 37.5 

(942 x 1986 x 953)

607

(275)

iUF118

208/230 V 

 

60 Hz

11.0

18

(510)

23.1 x 54.1 x 23.3 

(587 x 1374 x 592)

28.9 x 78.2 x 34.8 

(734 x 1986 x 884)

32.5 x 78.2 x 37.5 

(826 x 1986 x 953)

622

(282)

iUF124

208/230 V 

 

60 Hz

11.0

24 

(680)

34.4 x 49.5 x 23.3 

(874 x 1257 x 592)

45.1 x 78.2 x 34.8 

(1146 x 1986 x 884)

48.4 x 78.2 x 37.5 

(1229 x 1986 x 953)

704

(319)

iUF126

208/230 V 

 

60 Hz

11.0

26 

(736)

34.4 x 54.1 x 23.3 

(874 x 1374 x 592)

40.2 x 78.2 x 34.8 

(1021 x 1986 x 884)

43.8 x 78.2 x 37.5 

(1113 x 1986 x 953)

725

(328)

Содержание i.Series iUF116

Страница 1: ...Ultra Low Freezer Operation Manual i Series 360172 A D i Series iUF118 iUF126 i Series iUF116 iUF124...

Страница 2: ...tor with necessary instructions on the proper use and maintenance of certain Helmer Scientific products Any failure to follow the instructions as described could result in impaired product function in...

Страница 3: ...optional 7 3 i Series Operation 8 3 1 Initial Start Up 8 3 2 Operation 9 3 3 Change Temperature Setpoint 10 3 4 Set Alarm Parameters 11 3 5 Active Alarms 12 3 6 Mute and Disable Active Alarms 12 4 Mi...

Страница 4: ...eful information regarding a procedure or operating technique when using Helmer Scientific products NOTICE Advises the user against initiating an action or creating a situation which could result in d...

Страница 5: ...nts The freezer is not considered to be a storage cabinet for flammable or hazardous materials CAUTION Decontaminate parts prior to sending for service or repair Contact Helmer or your distributor for...

Страница 6: ...te 5 Insert metal spacer in the center of rubber compressor foot 6 Insert hex bolt through bottom of freezer and up through compressor foot and mounting plate 7 Place washer and hex nut over the protr...

Страница 7: ...and offs with the corresponding threaded holes on the back of the freezer 2 Insert the 3 8 hex head cap screws through the holes in the stand offs 3 Hand thread the cap screws into the threaded holes...

Страница 8: ...Install Replace Chart Paper Note For accurate temperature reading ensure current time is aligned with time line groove when the chart knob is fully tightened Chart recorder stylus and time line groov...

Страница 9: ...le 2 Switch battery ON OFF switch ON 3 Switch AC ON OFF switch ON Note The Start screen is displayed when the i C3 is powered on The i C3 monitoring and control system will take approximately three mi...

Страница 10: ...Temperature occurs refer to the service manual for troubleshooting 3 2 Operation Notes Refer to the i C3 User Guide for Ultra Low Freezers for complete information regarding the i C3 User Interface T...

Страница 11: ...point toggle button can be accessed from either the initial Settings screen or the Device Control Settings screen Enter the Settings password Select Temperature Setpoints Touch minus or plus on spin b...

Страница 12: ...Parameters Enter the Settings password Scroll down to select Alarm Settings Touch minus or plus on spin box to set each alarm parameter Settings screen Alarm Settings screens Alarm settings control th...

Страница 13: ...frigeration System Refrigerant pressure is too high High stage compressor temperature is too high Low stage compressor temperature is too high High stage compressor has failed Low stage compressor has...

Страница 14: ...ure reading since the last system reset power on event or manually initiated reset Touch the Reset icon to the right of the display to manually reset Notes The Min Max temperature display can be turne...

Страница 15: ...Ultra Low Freezers for complete information regarding Access Control 5 1 Setup Configure and manage multiple user specific accounts to allow controlled access to the freezer Access Setup Access Contr...

Страница 16: ...Operation Manual 360172 A D 15 5 2 Open Freezer with Access Control Access Control Keypad Using Access Control to unlock the door 1 Enter unique 4 digit PIN 2 If a valid PIN is entered the Padlock ind...

Страница 17: ...rm monitoring system is intended for connection to the end user s central alarm system s that uses normally open or normally closed dry contacts If an external power supply exceeding 30 V RMS or 60 V...

Страница 18: ...aled steel with bacteria resistant powder coated finish External Material Galvannealed steel with bacteria resistant powder coated finish External Port 2 standard top left corner rear of cabinet botto...

Страница 19: ...an 70 dB A 7 2 Electromagnetic Compliance This device is suitable for use in a specific electromagnetic environment The end user of this device is responsible for ensuring the device is used in compli...

Страница 20: ...lectrical components and wiring terminals in the electrical box for discoloration Contact Helmer Technical Service if any discoloration is found Inspect and clean the condenser filter Replace the i C3...

Страница 21: ...con Description Icon Description Home i C Applications Icon Transfer Save Event Log Settings Download Cancel Mute Temperature Graph Upload Back Arrow Reset Information Log Access Control Scroll Arrows...

Страница 22: ...N OFF switch B USB port K Condenser discharge grill C Door handle with key lock electronic lock and padlock hasp L RJ 45 Ethernet port D Condenser grill and filter media M USB port E Side panel refer...

Страница 23: ...d damage to the interior door hinges do not apply upward or downward force to the interior doors Label Description Label Description A Inner door F Shelf standard B Inner door hinge G Shelf clip C Inn...

Страница 24: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2017 Helmer Inc 360172 A D...

Отзывы: