7
Secondary Probe
The secondary probe is located in the lower right corner of the refrigerator.
Secondary probe
Fill Temperature Probe Bottle
Note
For each probe bottle, use approximately 4 oz. (120 mL) of product simulation solution (10:1 ratio of water to glycerin)
(Packet included in refrigerator box).
1. Remove all probes from bottle and remove bottle from bracket.
2.
Remove cap and fill with approximately 4 oz. (120 mL) of product simulation solution.
3. Secure cap on bottle and place in bracket.
4. Replace probes, immersing at least 2” (50 mm) in solution.
Install Additional Probe Through Top Port
Top access port
1. Peel back putty to expose port.
2. Insert probe through port into chamber.
3. Insert probe into bottle.
4. Replace putty, ensuring a tight seal.
Install Additional Probe Through Side Port (If Included)
1. Remove interior and exterior plugs to expose side access port.
2. Insert probe through port into chamber.
3. Insert probe into bottle.
4. Ensure port is tightly sealed using putty.