 Refrigerator - Upright

Operation Manual 

360372/B

19

6.4  Set Parameter Values

1.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to enter program mode. 

2.  The LED Display will show .C or .F to indicate Celsius or Fahrenheit.
3.  Press and release 

SEL

 button to scroll through the parameters.

4.  Once the desired parameter is selected, press and hold the 

SET

 button while pressing the 

Up

 or 

Down Arrow

 to select 

the desired value.

5.  Release 

SET

 button.  The new setting is saved.

6.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to exit program mode.

  Note

Contact Helmer Technical Service to set Hysteresis values.

Table 3.   Parameter Values

Parameter

Visual Indicator

Range

Default

Celsius or Fahrenheit

None

˚C, ˚F

˚C

High Temperature

MONITOR Lamp & HIGH Lamp

-40.0 to 25.0 (˚C) -40 to 77 (˚F)

5.5 ˚C

Low Temperature

MONITOR Lamp & LOW Lamp

-40.0 to 25.0 (˚C) -40 to 77 (˚F)

1.5 ˚C (HB); 2.0 ˚C (HLR & HPR)

Monitor Offset

MONITOR Lamp

-10.0 to 10.0 (˚C) -18 to 18 (˚F)

Varies

Control Offset

CONTROL Lamp

-10.0 to 10.0 (˚C) -18 to 18 (˚F)

Varies

Hysteresis

CONTROL Lamp

0.5 to 2.5 (˚C) 1 to 5 (˚F)

2.0˚C (111)*

 

1.0˚C (120, 125)

 

1.5˚C (245, 256)

 

*

 Hysteresis for 11 cu ft units prior to serial number 2044369 in low humidity mode = 0.8 

˚C

.

6.5  Set Temperature Units

  Note

If temperature units are changed, the temperature setpoints, offsets and alarm settings must be recalibrated.

1.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to enter program mode.

2. 

The LED Display will show ˚C or ˚F to indicate Celsius or Fahrenheit.

3.  Press and hold the 

SET

 button while pressing the 

Up

 or 

Down Arrow

 to select the desired temperature unit.

4.  Release 

SET

 button. The new setting is saved.

5.  Press and hold the 

Up

 and 

Down Arrows

 simultaneously for 3 seconds to exit program mode.

6.6  Temperature Calibration Offsets

Temperature calibration offsets indicate an acceptable margin of error between the actual temperature value and the desired 
temperature value.

Monitor Offset

 

Adjust if temperature displayed on the monitor is outside of your range when compared to a calibrated reference 
thermometer in the same probe bottle.

 

Value is factory-set to match a calibrated reference thermometer.

 

Refer to the service manual for instructions regarding changing the Monitor Offset.

Horizon Series Operation

Содержание Horizon series

Страница 1: ...Series iLR111 iLR120 iLR125 iLR245 iLR256 Horizon Series HLR111 HLR120 HLR125 HLR245 HLR256 Pharmacy i Series iPR111 iPR120 iPR125 iPR245 iPR256 Horizon Series HPR111 HPR120 HPR125 HPR245 HPR256 Blood Bank i Series iB111 iB120 iB125 iB245 iB256 Horizon Series HB111 HB120 HB125 HB245 HB256 ...

Страница 2: ...c and Helmer Disclaimer This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain Helmer Scientific products Any failure to follow the instructions as described could result in impaired product function injury to the operator or others or void applicable product warranties Helmer Scientific accepts no responsibility for liab...

Страница 3: ...ng 15 5 i Series Access Control Optional 16 5 1 Set Up 16 5 2 Open Refrigerator with Access Control 16 6 Horizon Series Operation 17 6 1 Initial Power Up 17 6 2 Display Minimum and Maximum Monitor Temperature Recordings 17 6 3 Change Temperature Setpoint 18 6 4 Set Parameter Values 19 6 5 Set Temperature Units 19 6 6 Temperature Calibration Offsets 19 6 7 Set Alarm Setpoints Parameters 20 6 8 Acti...

Страница 4: ... for the user Task Indicates procedures which need to be followed Note Provides useful information regarding a procedure or operating technique when using Helmer Scientific products NOTICE Advises the user against initiating an action or creating a situation which could result in damage to equipment personal injury is unlikely CAUTION Advises the user against initiating an action or creating a sit...

Страница 5: ...lmer refrigerators are intended for the storage of blood products and other medical and scientific products General Use Allow refrigerator to come to room temperature before switching power on During initial startup high temperature alarm may sound while refrigerator reaches operating temperature CAUTION Do not remove the cover from the condensate evaporator tray Initial Loading Allow the refriger...

Страница 6: ...frigerator is level Note Helmer recommends the use of leveling feet 2 3 Temperature Probes Probe bottle s along with a container of glycerin have been provided with this unit The glycerin is used to create a solution which simulates the product stored in the refrigerator The product simulation solution temperature reflects the product s temperature during normal operation Notes Temperature probes ...

Страница 7: ...e bottle from bracket 2 Remove cap and fill with approximately 4 oz 120 mL of product simulation solution 3 Secure cap on bottle and place in bracket 4 Replace probes immersing at least 2 50 mm in solution Install Additional Probe Through Top Port Top access port 1 Peel back putty to expose port 2 Insert probe through port into chamber 3 Insert probe into bottle 4 Replace putty ensuring a tight se...

Страница 8: ... well as voltage level remaining If full battery power is available back up power for the temperature chart recorder is available for up to 14 hours Note If chart recorder is operated on battery power the battery should be replaced to ensure the back up source has proper charge Prior to use Place probe in bottle with primary monitor probe Set up and Operation Access chart recorder by pressing and ...

Страница 9: ...en the knob can result in paper slipping and losing time 6 Press and hold C button When stylus begins to move right release button 7 Confirm stylus is marking on paper and stops at the correct temperature 8 Calibrate chart recorder to match primary temperature if needed and close recorder door Power Supply The temperature chart recorder uses AC power when the system is operating If AC power fails ...

Страница 10: ...d ON with a key switch The Start screen is displayed when the i C3 is powered on The i C3 will take approximately 2 5 minutes to boot up Start screen The language screen is displayed when the i C3 is powered on Use the Language screen to select the i C3 display language Language screens Note English is the default language If an alarm sounds temporarily mute the alarm by touching the Mute button H...

Страница 11: ...tivity the screensaver will be displayed To return to the Home screen touch the screensaver Home screen Home screensaver touch to return to Home screen 3 3 Change Temperature Setpoint Note The Temperature Setpoint toggle button can be accessed from either the initial Settings screen or the Device Control Settings screen Enter the Settings password Select Temperature Setpoints Touch minus or plus o...

Страница 12: ...Alarm Parameters Enter the Settings password Scroll down to select Alarm Settings Touch minus or plus on spin box to set each alarm parameter Settings screen Alarms screens Alarm settings control the circumstances and timing of alarm condition indicators displayed on the i C Home screen i Series Operation ...

Страница 13: ...operly Control Probe Failure Control probe not functioning properly Compressor High Temperature Compressor temperature reading is above high temperature alarm setpoint Compressor Probe Failure Compressor probe not functioning properly Power Failure Power to unit has been disrupted Door Open Door is open beyond user specified duration Low Battery Rechargeable battery voltage is low No Battery Batte...

Страница 14: ...n Home Temperature Graph Upload Save Event Log Alarm Test Access Control Cancel Mute Information Logs Access Log Back Arrow Reset Contact Information Contact Helmer Defrost Cycle Scroll Zoom Information Display Brightness Defrost Log Temperature Graph Forward Back i C Applications Icon Transfer Alarm Conditions Battery Power Settings Download Cancel Test units with serial numbers prior to 2044369 ...

Страница 15: ...ture reading since the last system reset power on event or manually initiated reset Touch the Reset icon to the right of the display to manually reset Notes The Min Max temperature display can be turned on or off through Display Settings Once the time reaches the maximum display of 999 hours and 60 minutes the message will display 999 60 but minimum and maximum temperatures will continue to be tra...

Страница 16: ... for refrigerator contents Refer to i C3 User Guide for complete information regarding Access Control 5 1 Set Up Configure and manage user specific accounts to allow controlled access to the refrigerator Access Set Up Access Control Setup password screen Access Control Setup screen Enter the supervisor PIN to set up Access Control and follow the on screen prompts to add delete or edit user informa...

Страница 17: ...ature probe Units which do not include the minimum and maximum recording feature will not display C or F when entering the program mode The minimum and maximum recording feature allows the user to view a minimum temperature occurrence and a maximum temperature occurrence within a given period of time The timer provides a time reference in which those temperatures occurred View minimum temperature ...

Страница 18: ...ings 1 Press and hold either the Up or Down Arrow button for 1 second or 2 While the display is flashing the HI or LO value press and hold the SET button for 1 second and listen for a single beep 3 While the display is flashing the elapsed time since last reset press and hold the SET button for 2 seconds CLr will be displayed followed by a series of 3 beeps to indicate exit back to the temperature...

Страница 19: ...8 to 18 F Varies Hysteresis CONTROL Lamp 0 5 to 2 5 C 1 to 5 F 2 0 C 111 1 0 C 120 125 1 5 C 245 256 Hysteresis for 11 cu ft units prior to serial number 2044369 in low humidity mode 0 8 C 6 5 Set Temperature Units Note If temperature units are changed the temperature setpoints offsets and alarm settings must be recalibrated 1 Press and hold the Up and Down Arrows simultaneously for 3 seconds to e...

Страница 20: ...etting is saved 6 Press and hold Up and Down Arrows simultaneously for 3 seconds to exit program mode 6 8 Active Alarms The controller displays temperature and alarm information Table 4 Horizon Series Active Alarms Alarm Visual Indicator Description High Temperature HIGH TEMP lamp flashes Chamber temperature reading is above high temperature alarm setpoint Low Temperature LOW TEMP lamp flashes Cha...

Страница 21: ...enter user codes Note The master code should be changed to prevent unauthorized user code setup Enter unique user codes for up to 100 users Each user code is stored with a specific record location number Keep a log of the location numbers and user codes with users names Add User Code 1 Enter the master code followed by the asterisk key 2 Press 1 to initiate user code programming function 3 Enter t...

Страница 22: ...2 6A 230V only quantity 2 7A 230V only quantity 2 7A 230V only quantity 2 Current Draw 7 0A 115V 60Hz 3 5A 230V 50Hz 3 5A 230V 60Hz 7 5A 115V 60Hz 4 2A 230V 50Hz 4 2A 230V 60Hz 7 5A 115V 60Hz 4 2A 230V 50Hz 4 2A 230V 60Hz 11 5A 115V 60Hz 6 0A 230V 50Hz 6 0A 230V 60Hz 11 5A 115V 60Hz 6 0A 230V 50Hz 6 0A 230V 60Hz Power Source Grounded outlet meeting national electric code NEC in the U S and local e...

Страница 23: ...26 470 214 230V 50 Hz 4 2 230 V 60 Hz 4 2 iLR125 115V 60 Hz 7 5 25 2 714 Upright Single hinged glass 4 29 5 x 79 5 x 38 5 750 x 2020 x 978 484 220 230V 50 Hz 4 2 230 V 60 Hz 4 2 HLR125 115V 60 Hz 7 5 25 2 714 Upright Single hinged glass 4 29 5 x 78 75 x 38 5 750 x 2001 x 978 481 219 230V 50 Hz 4 2 230 V 60 Hz 4 2 iLR245 115V 60 Hz 11 5 44 9 1271 Upright Double hinged glass 8 59 25 x 79 5 x 32 5 15...

Страница 24: ... 826 528 240 230V 50 Hz 4 2 230 V 60 Hz 4 2 iB125 115V 60 Hz 7 5 25 2 714 Upright Single hinged glass 7 29 5 x 79 5 x 38 5 750 x 2020 x 978 559 254 230V 50 Hz 4 2 230 V 60 Hz 4 2 HB125 115V 60 Hz 7 5 25 2 714 Upright Single hinged glass 7 29 5 x 78 75 x 38 5 750 x 2001 x 978 556 253 230V 50 Hz 4 2 230 V 60 Hz 4 2 iB245 115V 60 Hz 11 5 44 9 1271 Upright Double hinged glass 14 59 25 x 79 5 x 32 5 15...

Страница 25: ...22 237 230V 50 Hz 4 2 230 V 60 Hz 4 2 iPR125 115V 60 Hz 7 5 25 2 714 Upright Single hinged glass 1 6 29 5 x 79 5 x 38 5 750 x 2020 x 978 552 251 230V 50 Hz 4 2 230 V 60 Hz 4 2 HPR125 115V 60 Hz 7 5 25 2 714 Upright Single hinged glass 1 6 29 5 x 78 75 x 38 5 750 x 2001 x 978 549 250 230V 50 Hz 4 2 230 V 60 Hz 4 2 iPR245 115V 60 Hz 11 5 44 9 1271 Upright Double hinged glass 2 12 59 25 x 79 5 x 32 5...

Страница 26: ...B A EC REP Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherlands 9 2 WEEE Compliance The WEEE symbol right indicates this product falls under the scope of the WEEE Waste Electrical and Electronic Equipment directive When disposing of this product in countries affected by this directive Do not dispose of this product as unsorted municipal waste Collect this product separately Use the c...

Страница 27: ...for one year Refer to the Temperature Chart Recorder Operation and Service Manual Check the level of the solution in the probe bottles Refill or replace solution if necessary Examine the probe bottles and clean or replace if necessary Check the chamber lights and replace them if necessary Clean the condenser grill Clean the door gaskets interior and exterior of the refrigerator If applicable test ...

Страница 28: ...tor probe bottle excluding 111 models V RS 232 Com port optional H Caster W Back up battery switch I Drawer basket slide X Main power switch J Standard for adjusting storage components Y Circuit breakers 230V models K Magnetic lock Access control optional feature Z Condenser L Back up battery key switch Access control optional feature AA Compressor M Condenser grill BB Back up battery Monitoring a...

Страница 29: ...porator with fan guard Q Power cord E Primary Monitor probe bottle R Condensate evaporator F Drawer S Water evaporation tray G Caster T Remote alarm interface H Back up battery Access control optional feature U Main power switch I Keypad Access control optional feature V Circuit breakers 230V models J Back up battery key switch Access control optional feature W Condenser K Magnetic lock Access con...

Страница 30: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA Copyright 2018 Helmer Inc 360372 B ...

Отзывы: