81
Recommended separation distances between
Portable and mobile RF communications equipment and UnicLine S dental unit
The UnicLine S Dental Delivery System is intended for use in an electromagnetic environment in
wich radiated RF disturbances are controlled. The customer or the user of UnicLine S Dental
Delivery System can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
UnicLine S Dental Delivery System as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P
0.01
0.2
0.2
0.3
0.1
0.4
0.4
0.7
1
1.2
1.2
2.4
10
4.0
4.0
8.0
100
12.0
12.0
24.0