Adjustment of water pressure to Cup and Bowl.
The water pressure can be adjusted on the
adjustments screws (+) for more pressure (-) for
less pressure.
Cup
Bowl
Dismantling the glass bowl
The glass bowl can be removed for
cleaning. Remove mushroom from
drain.
Unscrew the bowl by turn it counter clockwise.
Then lift it off.
Mount the bowl again by following the instruction in the opposite way.
Do not exceed 65
C/149
F when cleaning the bowl.
7.1
Turnable fountain
When a turnable fountain is installed, the chair is automatically
locked when the fountain is turned to the patient side. The
fountain plug is connected to pin 4 and pin 5 of the P223 on the
console PCB. The switching point of the switches is adjusted by
loosening the screws that hold the switch, after which you adjust
and tighten the switch again.