Setup | Single steps for setup
Tap
File management
in the main menu
Open in the sequence
Internal
User
DataTransfer
The folder contains the file
MyFormat1.xml
Copy the file
MyFormat1.xml
to a transfer medium
Rename file
Edit the file in an XML editor or the computer's text editor
Copy the file from the transfer medium to the following folder
of the product:
Internal
►
User
►
DataTransfer
Use the
Switch-off
menu to shut down the product and then
restart it
The data format can be selected through the following path:
Settings
►
Interfaces
►
Data transfer
In order to prevent your data formats from being removed when the
firmware is updated, save each file under a separate name.
When the firmware is updated, the
MyFormat1
file in the
DataTransfer
folder is reset to the status it had when shipped. The file is
automatically recreated if it no longer exists. Other files in the
DataTransfer
folder are not affected by a firmware update.
XML schema of the MyFormat1.xml file
<configuration>
<base id="Settings">
<group id="General">
<group id="Format">
<group id="MyFormat1">
<element id="General" prefix=" " suffix=" " previousValues="false" writeLabel="true" writeUnit="true" writeTimestamp="false" newlineAfterTimestamp="false"/>
<element id="X" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="X MIN" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="X MAX" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="X RANGE" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Y" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Y MIN" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Y MAX" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Y RANGE" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Z" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Z MIN" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Z MAX" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Z RANGE" unit="mm" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Q" unit="deg" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Q MIN" unit="deg" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Q MAX" unit="deg" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
<element id="Q RANGE" unit="deg" base="10" factor="1" newline="false" prefix=" " suffix=" " decimalPlaces="3" digits="0" positiveSign="false"/>
</group>
</group>
</group>
</base>
<base id="version" major="0" minor="0" build="0"/>
</configuration>
1
3
2
4
6
5
Figure 26: Data format
MyFormat1.xml
1
Header
2
Name of the data format that appears in the
Settings
menu
3
General data format settings
4
Settings of the axes
5
Footer
6
End of the data format
8
HEIDENHAIN | GAGE-CHEK 2000 | Operating Instructions | 08/2020
147
Содержание GAGE-CHEK 2000 Series
Страница 1: ...GAGE CHEK 2000 Operating Instructions Evaluation Unit English en 08 2020 ...
Страница 12: ......
Страница 13: ...1 Fundamentals ...
Страница 22: ......
Страница 23: ...2 Safety ...
Страница 28: ......
Страница 29: ...3 Transport and storage ...
Страница 34: ......
Страница 35: ...4 Mounting ...
Страница 41: ...5 Installation ...
Страница 51: ...6 Basic operation ...
Страница 74: ......
Страница 75: ...7 Commissioning ...
Страница 119: ...8 Setup ...
Страница 154: ......
Страница 155: ...9 Measuring ...
Страница 169: ...10 File management ...
Страница 176: ......
Страница 177: ...11 Settings ...
Страница 210: ......
Страница 211: ...12 Servicing and maintenance ...
Страница 220: ......
Страница 221: ...13 What to do if ...
Страница 225: ...14 Removal and disposal ...
Страница 227: ...15 Specifications ...