HEBU reusable APC probe
25
EN
7 Preparation
The person in charge of preparatory treatment is responsible for ensuring that the treatment is duly
carried out using the relevant equipment, materials and personnel in the treatment facility and so
achieves the desired result. This necessitates validation and routine monitoring of the process used.
We urge you to take note of the national regulations dealing with instrument preparation.
7.1 Reutilization restrictions
Frequent repeat preparatory treatment has minimal effects on the product. The end of the product life
is normally determined by wear and damage due to use
7.2 Information on instrument preparation
◼
Use cleaning and/or disinfection agents with a pH-value within 9-10. Please observe
manufacturer instructions regarding dosage, exposure time and renewal of solutions.
◼
Do
not
use hard brushes (e.g. metal brushes or metal sponges) or coarse abrasive cleaners.
◼
Never leave instruments in cleaning or disinfection agents for longer than the specified time.
◼
Only used demineralized water for rinsing.
◼
Rinse and dry carefully through channels and pipes.
◼
Sensitive instruments must be cleaned in a storage or clamping fixture.
◼
Observe manufacturer instructions of cleaning
– and sterilizing equipment.
7.3 Preparation at the place of use
Directly after using remove coarse dirt of products and rinse out the working cannulas. Do not use
fixing agents or hot water (> 40°C), as this results in residues becoming fixed and can affect the
success of the subsequent cleaning operation
Dismantle and/or open products as far as possible. Within short time after use clean the products for
reducing a drying of the residues.
This enables an easier cleaning. If products come into contact with corroding medicines or cleaning
agents, wash these up with water immediately after use.
Longer drying times, e.g. for dry disposal are not validated and therefore not recommended.
The drying time during validation was 1 hour.