Operator’s Manual
8
RMLD-FR
Doc No. 16086 Rev A, RMLD-FR Operator Manual
© 2021 Heath Consultants Incorporated. All rights reserved.
1. Turn instrument OFF.
2. Unscrew the two captive
screws located on the bottom
of battery pack. Remove battery
from instrument.
3. Insert battery into the charger,
ensuring it is all the way down.
Optional: Fasten captive screws
into charger housing to prevent
battery from dislodging.
4. LED panel of charger illuminates
to show charging status of the
battery.
a. Green flashing = charging
b. Green solid = fully charged
c. No light = battery not present
or not connected properly
d. Red = fault
i. Position battery into the
other bay, if still red, then
try another battery.
ii. If the light is still red send
to repair or replace.
5. When fully charged, power up
the instrument to verify the
battery capacity on the screen.
Charging Procedure
DANGER: To reduce the risk of
ignition of a flammable or explosive
atmosphere, the battery must be
recharged only in a location known
to be non-hazardous.
WARNING: Only use the HEATH
supplied RMLD-FR battery charger
to recharge the instrument. Use
of any other charger may cause
severe damage to the battery or
electrical circuits.
CAUTION: To prevent damage to
the battery or electrical circuits,
always plug the charger into a
surge-protected outlet.
NOTICE:The prolonged storage of
battery inside or outside the instru-
ment can lead to battery chemistry
being irreversibly damaged leading
to permanent failure of the battery.
Remaining Battery Capacity
Remaining battery capacity
(in percent) is indicated in the
top right of the screen.
DANGER
WARNING
CAUTION
NOTICE
When storing the
instrument or battery
for more than a month:
- Charge battery to 40-50%
- Store at 60-70° F
Full charge in 2-3 hours
Charge in ambient
temperature above
50° F (10° C) to obtain
full battery capacity
a
b
c
d
a
b
c
d
Charging legend displayed on charger
White icon indicates
sufficient capacity
remaining
Red icon indicates low
capacity remaining/
time to charge