Interfaces
PW15iA
A4359-1.0
HBM: public
27
7
Interfaces
The ground reference for all the interface signals is
based on the supply voltage ground (GND). GND (0V) of
the excitation voltage must therefore also be connected,
but you must not connect GND to the shield. Use a sepa
rate line to connect the node digital ground to the power
supply GND (0V).
Use a shielded cable as the interface cable. The shield
should always be connected to the housing at both ends.
Important
Before installing
several
digital load cells into an installa
tion with a bus system, take the following into account:
The printed production number (type plate) is required for
setting up data communication. If the type plate can no
longer be seen after installation, the numbers of each
digital load cell should be noted beforehand. This enables
different addresses to be assigned during initial opera
tion.
Alternatively,
before
connection to the bus system, you
can connect each digital load cell
individually with a PC,
in order to set different addresses (see
ADR
command in
the online help).
7.1
RS‐485 4‐wire interfaces (UART)
The PW15iA comes supplied with an RS-485 interface.
Bit rates of 1200 to 115,200 baud can be set for the inter
face.
Either a single digital load cell can be connected via the
RS485 interface, or you can set up a bus system to con
Содержание PW15iA
Страница 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15iA...
Страница 35: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15iA Digitale W gezelle...
Страница 67: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15iA Peson num rique...
Страница 102: ...limination des d chets et protection de l environnement 36 A4359 1 0 HBM public PW15iA...
Страница 103: ...limination des d chets et protection de l environnement PW15iA A4359 1 0 HBM public 37...