Structure and mode of operation
PW15iA
A4359-1.0
HBM: public
17
Notice
Acids and all substances that release ions also attack
stainless steels and their seam welds.
The resultant corrosion can cause the transducer to fail.
If this is the case, you must provide appropriate means of
protection.
4.2
Deposits
Dust, dirt and other foreign matter must not be allowed to
accumulate sufficiently to divert some of the measuring
force onto the housing, thus distorting the measured
value (force shunt).
Do not use hard or pointed objects when cleaning the
gap between load application and housing, or between
mounting plate and housing.