Structure and mode of operation
PW15iA
A4359-1.0
HBM: public
13
S
activate an automatic zero on start up function,
S
activate an automatic zero tracking function,
S
linearize the scale curve with a third order polynomial,
S
activate various digital filters. Available filters include
those with cut-off frequencies below 1Hz, fast-settling
filters for dynamic measurements, notch filters and
mean value filters.
Use
MSV?
to read out the current measured value. The
format of the measured value (ASCII or binary) is set
with the
COF
command. You can also use the
COF
com
mand to activate automatic measurement output. The
measured values are transmitted in the following format,
subject to the
COF
command:
Output format
Input signal
Output when NOV=0
Output when
NOV>0
Binary, 2 chars. (INT)
0
…
maximum
capacity
0
…
20,000digits
1
…
NOV
Binary, 4 chars. (LONG)
0
…
maximum
capacity
0
…
5,120,000digits
1
…
NOV
ASCII
0
…
maximum
capacity
0
…
1,000,000digits
1
…
NOV
3.3
Adaptive interference suppression
Whatever the mode of operation, you can use the
ADF
command to activate automatic interference suppression
with adaptive filters. Interference frequencies are auto
matically found during measurement and suppressed by
comb filters and averaging. The maximum filter settling
time can be limited with the
TMA
command.
Содержание PW15iA
Страница 1: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15iA...
Страница 35: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15iA Digitale W gezelle...
Страница 67: ...Mounting Instructions Montageanleitung Notice de montage English Deutsch Fran ais PW15iA Peson num rique...
Страница 102: ...limination des d chets et protection de l environnement 36 A4359 1 0 HBM public PW15iA...
Страница 103: ...limination des d chets et protection de l environnement PW15iA A4359 1 0 HBM public 37...