Assembly Instructions M8095-EN 

 

M8095-EN   2020-09 

 

9/13

 

 

  

 

16.

  Check flow rate calibration: 

  After calibration, run the unit for at least 5 minutes at 40 °C in 

normal mode. 

  Close shut-off valve between feed and return lines. 

  Flow rate switches to 0 L/min and alarm indication 

‚Flow rate 

zero

’ is given. 

 

NOTE! 

The alarm message ‚Flow rate zero’ is delayed in 
time. 

  If this is not the case, repeat the flow rate calibration. 

  

 

ATTENTION! 

Risk of incorrect calibration! 

False calibrations can lead to faults witch the unit. 

Therefore: 
–   Check the calibrations. 

   

17.

  Acknowledge alarm and open shut-off valve. 

 

18.

  Check unit functions. 

 

 

19.

  Switch the unit off by press the 

 key. 

  

 

Содержание M8095-EN

Страница 1: ...ch M8095 EN 2020 09 1 13 2020 09 Page Assembly Instructions M8095 EN Spare part unit board GIF 51 O ID T25240 X T27401 X Contents Purpose 2 Precondition 2 Procedure 2 Product group Thermo 5 3 Product...

Страница 2: ...GER Danger to life caused by electric current Touching conductive parts causes a direct danger to life Therefore For all work on the electrical system for maintenance cleaning or repair work disconnec...

Страница 3: ...Cable routing Fig 2 Cable routing housing size 3 pump G_ L_ 6_ 8_ 4 Proceed as follows in order to fit the new GIF 51 A4 On oil unit insert the printed circuit board level measuring A 10 into new GIF...

Страница 4: ...s plug and switch on main switch NOTE The unit will maybe perform an automatic software update Fig 4 Input window configuration NOTICE If configuration is incomplete the warning text Wrong entry will...

Страница 5: ...figuration 10 Configure unit to Software Version SW51 2 1431 Service Configuration menu using the Support user profile NOTICE When configuring note the following Select product group according to tabl...

Страница 6: ...1 Check that the unit is free of pressure Pressure gauge rear of unit 12 Calibrating the unit NOTICE A quality check can be carried out if required During the automatic check the most important measur...

Страница 7: ...main line BP 2 Fig 11 Calibrate the pressure sensor 2 Read off the current Main line pressure under Display Actual value Set the parameter Pressure sensor 2 offset under Service Calibrating Pressure a...

Страница 8: ...calculation New Flow rate internal offset 5 current Flow rate Open shut off valve Fig 13 Calibrating the flow rate From software version SW51 1 0849B onwards Operate unit in regular mode at 40 C for...

Страница 9: ...ate switches to 0 L min and alarm indication Flow rate zero is given NOTE The alarm message Flow rate zero is delayed in time If this is not the case repeat the flow rate calibration ATTENTION Risk of...

Страница 10: ...Remove the printed circuit board level measuring A 10 and plug in on the new GIF 51 Install new GIF 51 and fasten it with screw Connect plug connections Connection main pump cable X 62 X 64 Voltage 20...

Страница 11: ...bsequently conduct a Reset complete Fig 17 Software version display 8 Checking the software version NOTICE You can verify the currently installed software version using the button on the basic screen...

Страница 12: ...control unit TR Treat 5 water treatment unit CL Clean 5 cleaning unit Fig 20 Input window configuration from Software Version SW51 2 1449 NOTICE As of software SW51 2 1449 the configuration is reques...

Страница 13: ...unit Therefore Check the calibrations 13 Check unit functions 14 Switch the unit off by press the key Parts list Pos Description O ID O ID T27401 T27401 1 T27401 2 T25240 T25240 1 Pcs Pcs Pcs Pcs Pcs...

Отзывы: