Measuring Range e.g.:
SET-Turn-Switch Pos.:
Limit Value:
HAUBER-Elektronik GmbH
11
Applying a rectangular signal (24 V DC / 0,5 Hz) at pin 4, the Typ 663 overall functionality can be
checked. The test voltage simulates an vibration amplitude, that exceeds the maximum Limit
Value step.
To check the relays switch functionality the adjusted Delay Time has to be considered.
Fig. 10: Self Check-Signal
14
Self Check
Example:
Limit Value Adjustment
13
Limit Value Adjustment
0...32 mm/s
16 mm/s
8
t (s)
U (V)
0
0
0
0
0
0
0
1
0,5
1
2
4
8
16
2
1
2
4
8
16
32
3
1,5
3
6
12
24
48
4
2
4
8
16
32
64
5
2,5
5
10
20
40
80
6
3
6
12
24
48
96
7
3,5
7
14
28
56
112
8
4
8
16
32
64
128
9
4,5
9
18
36
72
144
10
5
10
20
40
80
160
11
5,5
11
22
44
88
176
12
6
12
24
48
96
192
13
6,5
13
26
52
104
208
14
7
14
28
56
112
224
15
7,5
15
30
60
120
240
SET-
Turn-Switch
Position
Limit Value
Range
0…8 mm/s
Range
0…16 mm/s
Range
0…32 mm/s
Range
0…64 mm/s
Range
0…128 mm/s
Range
0…256 mm/s
24 V DC