5.3 Waveform Generator (cont’d)
5.3.3 Experiment Delivery Screen
Functions and Operation (cont’d)
4.
If the researcher determines that the system voltage is
correct, pulsing is initialized by touching START or
depressing the foot pedal. The START button turns a
solid blue while pulsing is active and solid green when
pulsing is completed.
7.
Press the color-filled DONE button to return to the
opening screen.
5.
When the system reaches a shutdown state either by
successful waveform delivery or by error, all buttons will
turn solid green and pulse delivery data will be
displayed in the System message text box. The load
message box will change from the estimated load value
to the actual high voltage pulse monitored value.
6.
All run history log data are automatically stored in
internal Flash Memory as well as a USB Key, if one is
inserted.
Figure 5-8: Delivery Screen - High Voltage Power Supply Charge
Figure 5-9: Delivery Screen - Pulse Waveform Active
Figure 5-10: Delivery Screen - Shutdown