10
4.8 Single collect Assembly
Fig. 4.8.1
Setting up
Holder for Single
Cannula
Fig. 4.8.2
Setting up
Single Cannula for
Tubing
Fig. 4.8.3
Setting up Single
Cannula for Septa
4.8.1
Place the Holder for Single Cannula (1) so the tap (2) fit into the loop (3)
in the right side slot (4) on the Supporting Arm. Secure the holder with the two
accompanying screws (5). See Figure 4.8.1
4.8.2
Collecting in vials with tubing:
Insert the Cannula for Tubing in the Multi Holder for Cannula (6), adjust the
upper end of the metal cannula to the same level as the top of the Plastic
Locker Knob (7) when it is mounted in the Multi Holder for Cannula. Secure
the cannula with the screw in the Multi Holder for Cannula (8). See Figure
4.8.2
4.8.3
Collecting in vials with septa:
Place a waste vial in position 1 in the cassette and move it to the collection
position. Insert the Cannula for Septa (9) in the Multi Holder for Cannula (6).
Adjust the lower end of the cannula into the bottom of the vial (10). Secure
the cannula with the screw in the Multi Holder for Cannula (8). See Figure
4.8.3
4.8.4
To remove the Holder for Single Cannula, untight the two screws, turn
the Multi Holder for cannula so the opening of the locking washer (11), seen
under the supporting arm, is facing you. Lift up the Holder for Single Cannula
from the Supporting Arm. See Figure 4.8.1
2
3
4
5
5
6
7
8
6
8
7
9
10
11
1