10 TROUBLESHOOTING
If the CMA 470 Refrigerated Fraction Collector does not function, check that:
the input voltage and frequency correspond to the information given on the
nameplate on the bottom of the instrument.
the wall socket is live.
the power cables are undamaged and properly connected both to the wall
socket and to the instrument.
11 TECHNICAL DATA
Volume per Collection: 1 µL- 1,2 mL/min
Number of vials: 64 x 300µL, 40 x 2mL
Septa: Non reclosing (Teflon)
Cooling: Down to +6 °C in steps of 1 °C
Cooling capacity: -15 °C from environmental temperature or better
Temperature accuracy: +- 1.5 °C
Collection mode: Time, minutes and seconds
Environment temperature: 10 – 32 °C
Display: LCD 4 x 20 characters, back lit
Memory: Non - volatile. Stores all settings
External connections: General purpose digital Input/Output
Computer connection: RS 232 serial interface and USB (for user software)
Voltage: 100 – 240 VAC, 50 – 60 Hz
Power consumption: External adaptor 12 VDC, 55 W
Dimensions: 222 x 279 x 142 (167) mm (W x D x H)
Weight: Approx. 3.8 kg
Standards:
EN 61010-1 and IEC 61326
Directives:
LVD (Low Voltage, 73/23/EEC and 93/68/EEC),
EMC (2004/108/EG)
CMA 470 is designed to operate in a controlled electromagnetic environment,
i.e. where radio frequency transmitters such as mobile telephones may not be
used in close proximity.
Intended use: Designed for research and industrial applications, the CMA 470
Refrigerated Fraction Collector is not approved for clinical use.