HI 6600 Series Modular Sensor System User Guide
Page | 75
On the HI 6110 optional display
: select the communication menu and use the UP or DOWN button to
select the Modbus-RTU submenu, then select the Baud Rate menu item. The current Baud Rate value will be
displayed, press the ENTER or the RIGHT button if the value needs to be modified.
Parity Parameter
The Parity parameter has three options, EVEN, ODD, or OFF.
On the Web page:
select the communication menu and then select the Modbus RTU submenu, and left
click inside the Parity field and select even, odd or off using the drop down menu.
On the
HI 6110 optional display
, select the communication menu and use the UP or DOWN button to select
the Modbus-RTU submenu, then select the Slave Address menu item. The current Parity value will be
displayed, press the ENTER or the RIGHT button if the Parity needs to be modified.
1)
Setup a slave address assigned to the HI 6600 series instrument from the communications menu and the
Modbus-RTU submenu.. Use a unique address between 1 and 247
2)
Set the Baud Rate to the same as the master device, typically 9600 or 19,200 Step 3. Set the Parity Bit
parameter to match that of the Modbus-RTU master (EVEN,ODD, or OFF)
The RS 485 half duplex pinout is:
GND SHIELD
TX+
TX-
Signal
Ground
Do Not
Connect
TXD+ /RXD+
Non-inverting
TXD-/RxD-
Inverting
Содержание HI 6110
Страница 16: ...HI 6600 Series Modular Sensor System User Guide Page 16 ...
Страница 44: ...HI 6600 Series Modular Sensor System User Guide Page 44 ...
Страница 96: ...HI 6600 Series Modular Sensor System User Guide Page 96 General Troubleshooting Flow Chart Index ...
Страница 97: ...HI 6600 Series Modular Sensor System User Guide Page 97 A ElectricalandMechanicalReview ...
Страница 98: ...HI 6600 Series Modular Sensor System User Guide Page 98 A1 CheckingforUnstableComponentsinaWeighingSystem ...
Страница 100: ...HI 6600 Series Modular Sensor System User Guide Page 100 B1 GuidelinestoVerifyElectricalInstallation ...
Страница 101: ...HI 6600 Series Modular Sensor System User Guide Page 101 B2 GuidelinestoVerifyMechanicalInstallation ...
Страница 102: ...HI 6600 Series Modular Sensor System User Guide Page 102 B3 VerifyConfiguration FilterSettingstoImproveStability ...
Страница 103: ...HI 6600 Series Modular Sensor System User Guide Page 103 C IntegratedTechnicianandStabilityTestOverview ...
Страница 104: ...HI 6600 Series Modular Sensor System User Guide Page 104 ENon ReturntoZero SystemwithITSummingCard ...
Страница 106: ...HI 6600 Series Modular Sensor System User Guide Page 106 G Calibration Errors During Calibration ...
Страница 107: ...HI 6600 Series Modular Sensor System User Guide Page 107 H Mechanical Installation ...
Страница 108: ...HI 6600 Series Modular Sensor System User Guide Page 108 J Electrical Inspection ...
Страница 109: ...HI 6600 Series Modular Sensor System User Guide Page 109 K InstallationCheckPoints ...
Страница 110: ...HI 6600 Series Modular Sensor System User Guide Page 110 M WeightProcessor sOptionalFrontDisplayBlankorLocked ...
Страница 128: ...HI 6600 Series Modular Sensor System User Guide Page 128 ...
Страница 130: ...HI 6600 Series Modular Sensor System User Guide Page 130 Image of HI 6600 Hardy Gateway Module II Diagram ...
Страница 131: ...HI 6600 Series Modular Sensor System User Guide Page 131 Image of HI 6610 Weight Processing Module II Diagram ...
Страница 132: ...HI 6600 Series Modular Sensor System User Guide Page 132 OptionalHI6110DisplayPanelMountingTemplate ...