HI 6600 Series Modular Sensor System User Guide
Page | 45
C2 Calibration Process
Navigate to the C2 and eCAL menu from the Web brower or on the front panel display
1)
Check to make sure the number of load cells displayed in the Num Devices menu corresponds to
the actual number of C2 devices installed. If the number varies, check to ensure that each
load cell/point cable connection is securely fastened and that each load cell/point cable is not
broken. Only Hardy Process Solutions load sensors are C2 (or eCal) calibration capable.
2)
In the “Ref Weight” text field, enter the reference weight for your application. An empty scale
will use a 0.00 reference weight setting.
3)
Enter the Gravity Correction Factor, see the table above, then enter the Correction Factor
number you selected from the table in the Gravity Correction field that best corresponds to
your location.
4)
Click the “Do C2 Calibration” button.
Wait a few seconds and the results will appear.
If the Calibration was successful a Cal completed OK appears and gross weight will read your reference
value.
If the Calibration failed, check to make sure C2 load cells are being use and properly connected.
If the system is using one or more non-C2 load cells, calibrate using a traditional calibration (Hard Cal)
process.
See the Troubleshooting section of this manual for further information.
5)
Click “back” in the Web browser or use the left arrow key on the front panel display to return to the main
Calibration menus.
Содержание HI 6110
Страница 16: ...HI 6600 Series Modular Sensor System User Guide Page 16 ...
Страница 44: ...HI 6600 Series Modular Sensor System User Guide Page 44 ...
Страница 96: ...HI 6600 Series Modular Sensor System User Guide Page 96 General Troubleshooting Flow Chart Index ...
Страница 97: ...HI 6600 Series Modular Sensor System User Guide Page 97 A ElectricalandMechanicalReview ...
Страница 98: ...HI 6600 Series Modular Sensor System User Guide Page 98 A1 CheckingforUnstableComponentsinaWeighingSystem ...
Страница 100: ...HI 6600 Series Modular Sensor System User Guide Page 100 B1 GuidelinestoVerifyElectricalInstallation ...
Страница 101: ...HI 6600 Series Modular Sensor System User Guide Page 101 B2 GuidelinestoVerifyMechanicalInstallation ...
Страница 102: ...HI 6600 Series Modular Sensor System User Guide Page 102 B3 VerifyConfiguration FilterSettingstoImproveStability ...
Страница 103: ...HI 6600 Series Modular Sensor System User Guide Page 103 C IntegratedTechnicianandStabilityTestOverview ...
Страница 104: ...HI 6600 Series Modular Sensor System User Guide Page 104 ENon ReturntoZero SystemwithITSummingCard ...
Страница 106: ...HI 6600 Series Modular Sensor System User Guide Page 106 G Calibration Errors During Calibration ...
Страница 107: ...HI 6600 Series Modular Sensor System User Guide Page 107 H Mechanical Installation ...
Страница 108: ...HI 6600 Series Modular Sensor System User Guide Page 108 J Electrical Inspection ...
Страница 109: ...HI 6600 Series Modular Sensor System User Guide Page 109 K InstallationCheckPoints ...
Страница 110: ...HI 6600 Series Modular Sensor System User Guide Page 110 M WeightProcessor sOptionalFrontDisplayBlankorLocked ...
Страница 128: ...HI 6600 Series Modular Sensor System User Guide Page 128 ...
Страница 130: ...HI 6600 Series Modular Sensor System User Guide Page 130 Image of HI 6600 Hardy Gateway Module II Diagram ...
Страница 131: ...HI 6600 Series Modular Sensor System User Guide Page 131 Image of HI 6610 Weight Processing Module II Diagram ...
Страница 132: ...HI 6600 Series Modular Sensor System User Guide Page 132 OptionalHI6110DisplayPanelMountingTemplate ...