14
Configure Settings
1
MARKERS
: Colony count markers can be customized to plate type. Users have the
option to change the color, shape, weight, and size of the markers for different
plate modes.
2
SAVE IMAGE/SAVE DATA ONLY
: When this option is highlighted green, saved files
will include an image of the plate with the corresponding plate read data. When
option is in gray, saved files will include only the raw data from the plate read.
3
SAVE MARKERS ON/OFF
: When this option highlighted green for ON, saved files
will include two images, one with colony markers and the original raw image of the
plate. When option is gray, saved file will only include raw image of plate.
4
USB IMAGE
: Allows users to review and read plates stored on a USB drive.
5
LIVE IMAGE
: The default setting for the Wizard reader allows users to read
CompactDry™ plates in real time.
6
ADVANCED CONFIGURATION
: This option will be used when calibrating and
validating The Wizard™ CompactDry™ Reader for your sample matrix.
7
MENU
: Takes users back to plate mode configuration screen.
2
3
1
6
7
4
5
Содержание The Wizard CompactDry Reader