E. Disintegration Testing
Phase One Disintegration Tester User Guide—40-108-001 Rev. A—14 November 2016
E-7
Free Run Test
To set up a free-run (untimed, continuous) test, at the main
menu highlight FREE RUN but do not yet press the control
knob.
Before activating this option, ensure the media is up to
temperature, the baskets and dosage are in position, and any
other test preparations are complete. Use the TOOLS MENU as
needed to help verify the setup.
With all preparations completed and FREE RUN highlighted,
press the control knob to start the test. Note that the arms will
begin moving immediately.
At this point the screen will display the elapsed time (counting
UP from zero), the number of reciprocating cycles completed so
far, and the PAUSE function will be highlighted.
Press the control knob at any time to pause the test. For further
details on this function, see Pausing a Test.