8
SPECIFICA
TIONS
Resistivity
Range
0 to 999999 Ω·cm; 0 to 1000.0 kΩ·cm; 0 to 1.0000 MΩ·cm
Resolution Depending on resistivity reading
Calibration Based on conductivity calibration
Salinity
Range
0.00 to 70.00 PSU
Resolution 0.01 PSU
Accuracy
±2% of reading or ±0.01 PSU whichever is greater
Calibration Based on conductivity calibration
Seawater
Sigma
Range
0.0 to 50.0
σ
t
,
σ
0
,
σ
15
Resolution 0.1
σ
t
,
σ
0
,
σ
15
Accuracy
±1.0
σ
t
,
σ
0
,
σ
15
Calibration Based on conductivity or salinity calibration
Atmospheric
Pressure
Range
450.0 to 850.0 mmHg; 17.72 to 33.46 inHg;
600.0 to 1133.2 mbar; 8.702 to 16.436 psi;
0.5921 to 1.1184 atm; 60.00 to 113.32 kPa
Resolution 0.1 mmHg; 0.01 inHg; 0.1 mbar; 0.001 psi; 0.0001 atm; 0.01 kPa
Accuracy
±3.0 mmHg within ±15 °C from calibration temperature
Calibration Automatic at 1 custom point
Temperature
Range
-5.00 to 55.00 °C; 23.00 to 131.00 °F; 268.15 to 328.15 K
Resolution 0.01 °C; 0.01 °F; 0.01 K
Accuracy
± 0.15 °C; ± 0.27 °F; ±0.15 K
Calibration Automatic at 1 custom point
Temperature Compensation Automatic from -5 to 55 °C (23 to 131 °F)
Logging Memory
45,000 records (continuous logging or log-on-demand of all parameters)
Logging Interval
1 second to 3 hours
PC Interface
HI9298194
software
Protection Rating
IP67 (water-resistant)
Environment
0 to 50 °C (32 to 122 °F); RH 100 %
Battery Type
1.5V, AA alkaline batteries (4 pcs.)
Battery Life
360 hours of continuous use without backlight
Dimensions
185 x 93 x 35.2 mm (7.3 x 3.6 x 1.4”)
Weight
400 g (14.2 oz.)