38
SYSTEM SETUP
Auto Poweroff
Option: not used, 5, 10, 15, 20, 30, 60 minutes
The function is used to save battery life. After the set time is elapsed, the meter will:
1. Automatically switch off, if no key is pressed in normal measurement mode. Press
On/Off
key
to switch on again.
2. Enter in power save mode, if continuous logging mode is selected with a minimum logging
interval of 30 seconds. The “Power save mode” message and the
Wake up
functional key will
be displayed on the LCD when it goes into this mode. Logging is not stopped. Pressing
Wake up
will exit the Power save mode.
Key Beep
Option: Enabled or Disabled
When enabled, an acoustic signal is heard every time a key is pressed.
Press the functional key to select the desired option.
Error Beep
Option: Enabled or Disabled
Short beep: pressed key is incorrect. Long beep: pressed key is not active or an error is detected.
Press the functional key to select the desired option.
Decimal Separator
Option: Comma (,) or Period (.)
Select decimal separator used for display and log files.