46
LOGGING
12.2. START METER LOG
1. Select “Start meter log” to log the currently enabled parameters at the set logging interval on
the meter.
2. To set the logging interval, press
Options
. The log interval time can set from 1 second to
3 hours. Press
Modify
and use the alphanumeric keys to enter the desired log interval. Press
Accept
to confirm.
3. Press
Select
to to edit the lot, remark or tag.
4. To stop the meter log, enter the log menu and select “Stop meter log”.
12.3. LOG RECALL
Select “Log recall” to view or delete logs that are stored on the meter. The number of available lots
is displayed.
Lots
Select this option to view or delete individual lots.
1. Use the arrow keys to select the desired lot and press
View
. The meter displays a summary of
all data related to the selected lot: number of samples, memory space used, time and date of
the first and last readings.
2. Press
View
to review the log data. Use the arrow keys to change the sample number in the
selected lot. The sample number is shown on the bottom right corner of the display.
Notes:
The values displayed for compensated conductivity and TDS are dependent on the
coefficients defined in Parameter Coefficients in Setup menu (EC Reference Temperature, EC
Temperature Coefficient, TDS factor and Salinity).
If parameter units or coefficients are altered, stored logs on this meter will be altered reflecting
these changes. Save logs to a PC before altering parameters or coefficients.
Details are available only for the enabled parameters.
3. Press
Info
to see record information for the current sample (time & date, remark or serial
number, if available).