- When the water is good for irrigation, the Feed OK LED is ON.
- If any alarm condition occurs (see pag.22), the ALARM LED turns ON.

The controller can be restarted by pressing the
RESET button.

If the alarm condition is still present, the controller goes back in alarm
status.
When the controller is powered, the actual pH and conductivity values
are displayed on the two LCDs in pH and mS/cm units respectively.

NORMAL OPERATION

When the 

HI 9914

 controller is turned on, the meter performs a status

check of the level sensors and provides the commands depending on the
input status.
All the inputs have a delay of about 1 second in order to avoid wrong
commutations caused by the waves.

Step 1

:

At start-up the controller checks the normal level and low-level sensors.
• If the tank is empty, both sensors are open and the controller goes

to step 2.

• If  the low-level sensor is closed and the normal level sensor is open (the

water level is between low and normal level), the controller goes to
step 3.

Step 2

:

The water nozzle is activated (open) and the tank is filled with water.
When the low level is reached, the controller goes to step 3.
• If  the controller remains in this status for more than 30 seconds, an

alarm condition occurs.

Step 3

:

The water nozzle is open, the circulation pump is running and the tank
is filled with water.
When the normal level is reached (the normal level sensor is closed) the
controller goes to step 4.

Main power supply connection

Before connecting the unit to the mains, make sure the controller is
completely wired and all the connections for pumps, alarm, probes, etc.
have been done.
• Connect the mains power wires to the power

supply input pins.

• Replace the right panel lid and tighten the four

provided screws.

• Connect the controller to the mains.

CONTROLLER START-UP

Before starting-up make sure the controller has been properly calibrated,
and the pH and conductivity setpoints have been adjusted (see the
following pages).
The pH electrode, conductivity probe and any reference pin must be
properly connected and wired to the controller (see the section above).
The EC probe must be immersed in the solution above the air-vent holes
on the external sleeve, and installed in such a way to minimize the
presence of air bubbles (see probe installation tips at the end of the
manual).
Remove the protective cap from the pH electrode and immerse the
electrode (at least 4cm/1.5") into the solution to be tested. Install the
electrode in such a way that it permanently lies in the solution.
The level sensors, user input, external fill button and humidity sensor
must be connected accordingly with the chosen tank assembly.
If some of the inputs are not used, they must be connected as explained
in the TANK ASSEMBLY section.
The pumps, water nozzle and alarm must be wired as described above.
Turn on the water supply and start the controller by connecting it to the
mains power supply.
The 

HI 9914

 controls different status and gives visual information through

various LED:
- When the dosing pumps are active, the corresponding EC and pH pump
LEDs are on.
- When the circulation or the feeding pump is running, the corresponding
LED is ON.
- When the water nozzle is open, the water nozzle LED is ON.

Содержание HI 9914

Страница 1: ...1 Instruction Manual HI 9914 Wall mounted Fertigation Controller w w w h a n n a i n s t c o m...

Страница 2: ...y on two large LCDs with backlight feature for a best easy reading TheECprobeisdesignedwithabuilt intemperaturesensorwhichallows the controller to automatically compensate for the temperature effect T...

Страница 3: ...erdosing of fertilizer or acid The controller is designed with relay outputs for pH and conductivity control 2A 240V each one corresponding to a status LED for visual check Three level sensors are use...

Страница 4: ...eeding pump output 20 Circulation pump output 21 Alarm output 1 LCD for Conductivity readings 2 EC SET button 3 EC slope adjustment trimmer 4 EC MEAsure button 5 Alarm LED 6 Reset button 7 LCD for pH...

Страница 5: ...djustable from 0 5 to 14 0 pH Calibration Manual 2 point with offset 2 pH and slope 80 to 120 trimmers Analogoutput 0 7V 5 0 5V pH Controller Output 2A 220V relay Other Features Timer Adjustable from...

Страница 6: ...erenttankvolumesandwithvariouscomplex ity of tank assembly For proper operation of the controller the tank assemblies must be set minding the following the pH electrode and EC probe should rapidly det...

Страница 7: ...less of the water composition or level The controllerwillrefillthetankifthewatergoesundernormallevelforabout 15 minutes or if the fill button is pressed Note The inlet of the circulation pump must be...

Страница 8: ...the NC pin when is on the COMMON pin is connected with the NO pin The NC pin is not accessible for the pH and EC pumps and the COMMON Place the low level sensor to indicate the minimum water level fo...

Страница 9: ...onnect one pin of the source to the COMMON pin of the alarm output Connect the alarm device between the NO output pin and the remaining source pin Acid and fertilizer dosing pump connection The pH and...

Страница 10: ...ke sure the controller is completely wired and all the connections for pumps alarm probes etc have been done Connect the mains power wires to the power supply input pins Replace the right panel lid an...

Страница 11: ...e conductivity and pH values are good for at least 30 seconds the controller goes to step 5 If the normal level sensor remains open for about 15 minutes the controller goes back to step 3 Step 5 The C...

Страница 12: ...is reached SETTINGTHEECCONTROL To adjust the Conductivity setpoint Push the SET button close to EC display to read the setpoint value ALARM CONDITIONS When an alarm condition is reached the ALARM LED...

Страница 13: ...swillbeactivefor 3 minutes and off for 12 minutes 3 12 15 minutes Setting the timer In order to activate the timed dosing feature put the TIMED CONT switch to the TIMED position To set the duty cycle...

Страница 14: ...25 C Immerse the probe in the beaker ensuring that the holes on the probe sleevearecompletelycovered CALIBRATION Disconnect the pumps the water nozzle and the alarm or assure that the start of one of...

Страница 15: ...trode stem or bulb If any scratches or cracks are present replace the electrode The connector must be perfectly clean and dry Cleaning procedure Soak the electrode in HI 7061 General Cleaning Solution...

Страница 16: ...perly or constantly reads zero or a value close to it Check the probe and cable for cracks and replace it if needed If the controller display shows 1 The cable may be shorted or the probe broken Repla...

Страница 17: ...ottle EC calibration solutions HI7031L 1 41 mS cm calibration solution 500 mL bottle HI7039L 5 00 mS cm calibration solution 500 mL bottle Maintenancesolutions HI70300L Electrode storage solution 500...

Страница 18: ...ystem utilized in nursery and cutting plants cultivation Reference Principitecnico agronomicidellafertirrigazioneedelfuorisuolo pag 115 publishedby VenetoAgricoltura CentroSperimentaleOrtofloricoloPod...

Страница 19: ...37 36 APPENDIX C INSTALLATIONEXAMPLES Block diagram for a typical installation of an irrigation system with HI9914controller Block diagram for an irrigation system with recirculation circuit...

Страница 20: ...ribed main tenance are not covered This warranty is limited to free of charge or replacement of the meter only if any malfunctioning is due to manufac turing defects If service is required contact the...

Страница 21: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax...

Отзывы: