CONDUCTIVITY PROBE MAINTENANCE

Preparation

Make sure that the protective sleeve is on the probe shaft and is intact.

Storage

Conductivity probes should be stored dry. If they are not to be used for a
while, clean and dry them thoroughly before storing them in a dry place.

Periodic maintenance

Inspect the probe and the cable. The cable used for the connection to the
controller must be intact. There must be no points of broken insulation on
the cable or cracks on the probe sleeve. If any cracks are present, replace
the probe and cable.
The connector must be clean and dry.

Cleaning procedure

Soak the probe in 

HI 7061

 General Cleaning Solution for 1 hour.

If the probe has been left in highly concentrated fertilizer solution and
does not seem to have become clean, repeat the cleaning procedure. The
pins can also be cleaned with a cloth. The cloth has to be made of a soft
and nonabrasive material and does not scratch the pins.

Note: 

After performing a the cleaning procedure, rinse the probe thor-

oughly with distilled or tap water. Dry the probe and recalibrate
the controller.

Troubleshooting

• If the controller does not respond properly or constantly reads zero or
a value close to it:
- Check the probe and cable for cracks and replace it if needed.
• If the controller display shows 

1.

 :

- The cable may be shorted or the probe broken. Replace it if needed.
• If the response seems sluggish:
- Follow the above cleaning procedure.
• If there are anomalies such as fluctuating numbers:
- Ensure that the probe has been properly mounted and constantly lies

in a well filled with solution.

Troubleshooting

Evaluate  the electrode performance based on the following:
• Noise (readings fluctuate up and down) could be due to clogged/dirty

junction:

- Dry Membrane/Junction: soak in 

HI 70300

 Storage Solution overnight.

Check that the installation has been done to create a well to maintain
the electrode bulb constantly moist.
• Low Slope:
- Check the electrode for cracks in the glass stem or bulb (replace the
electrode if cracks are found).
- Make sure that cable and connections are neither damaged nor lying
in a pool of water or solution.
• Slow Response/Excessive Drift:
- Soak the tip in 

HI 7061

 Solution for 30 minutes, rinse thoroughly with

distilled water and then recalibrate the meter.

Note:

 It is always recommended to keep at least one spare electrode

handy. When anomalies are not resolved with a simple mainte-
nance, change the electrode (and recalibrate the controller) to see
if the problem is solved.

Содержание HI 9914

Страница 1: ...1 Instruction Manual HI 9914 Wall mounted Fertigation Controller w w w h a n n a i n s t c o m...

Страница 2: ...y on two large LCDs with backlight feature for a best easy reading TheECprobeisdesignedwithabuilt intemperaturesensorwhichallows the controller to automatically compensate for the temperature effect T...

Страница 3: ...erdosing of fertilizer or acid The controller is designed with relay outputs for pH and conductivity control 2A 240V each one corresponding to a status LED for visual check Three level sensors are use...

Страница 4: ...eeding pump output 20 Circulation pump output 21 Alarm output 1 LCD for Conductivity readings 2 EC SET button 3 EC slope adjustment trimmer 4 EC MEAsure button 5 Alarm LED 6 Reset button 7 LCD for pH...

Страница 5: ...djustable from 0 5 to 14 0 pH Calibration Manual 2 point with offset 2 pH and slope 80 to 120 trimmers Analogoutput 0 7V 5 0 5V pH Controller Output 2A 220V relay Other Features Timer Adjustable from...

Страница 6: ...erenttankvolumesandwithvariouscomplex ity of tank assembly For proper operation of the controller the tank assemblies must be set minding the following the pH electrode and EC probe should rapidly det...

Страница 7: ...less of the water composition or level The controllerwillrefillthetankifthewatergoesundernormallevelforabout 15 minutes or if the fill button is pressed Note The inlet of the circulation pump must be...

Страница 8: ...the NC pin when is on the COMMON pin is connected with the NO pin The NC pin is not accessible for the pH and EC pumps and the COMMON Place the low level sensor to indicate the minimum water level fo...

Страница 9: ...onnect one pin of the source to the COMMON pin of the alarm output Connect the alarm device between the NO output pin and the remaining source pin Acid and fertilizer dosing pump connection The pH and...

Страница 10: ...ke sure the controller is completely wired and all the connections for pumps alarm probes etc have been done Connect the mains power wires to the power supply input pins Replace the right panel lid an...

Страница 11: ...e conductivity and pH values are good for at least 30 seconds the controller goes to step 5 If the normal level sensor remains open for about 15 minutes the controller goes back to step 3 Step 5 The C...

Страница 12: ...is reached SETTINGTHEECCONTROL To adjust the Conductivity setpoint Push the SET button close to EC display to read the setpoint value ALARM CONDITIONS When an alarm condition is reached the ALARM LED...

Страница 13: ...swillbeactivefor 3 minutes and off for 12 minutes 3 12 15 minutes Setting the timer In order to activate the timed dosing feature put the TIMED CONT switch to the TIMED position To set the duty cycle...

Страница 14: ...25 C Immerse the probe in the beaker ensuring that the holes on the probe sleevearecompletelycovered CALIBRATION Disconnect the pumps the water nozzle and the alarm or assure that the start of one of...

Страница 15: ...trode stem or bulb If any scratches or cracks are present replace the electrode The connector must be perfectly clean and dry Cleaning procedure Soak the electrode in HI 7061 General Cleaning Solution...

Страница 16: ...perly or constantly reads zero or a value close to it Check the probe and cable for cracks and replace it if needed If the controller display shows 1 The cable may be shorted or the probe broken Repla...

Страница 17: ...ottle EC calibration solutions HI7031L 1 41 mS cm calibration solution 500 mL bottle HI7039L 5 00 mS cm calibration solution 500 mL bottle Maintenancesolutions HI70300L Electrode storage solution 500...

Страница 18: ...ystem utilized in nursery and cutting plants cultivation Reference Principitecnico agronomicidellafertirrigazioneedelfuorisuolo pag 115 publishedby VenetoAgricoltura CentroSperimentaleOrtofloricoloPod...

Страница 19: ...37 36 APPENDIX C INSTALLATIONEXAMPLES Block diagram for a typical installation of an irrigation system with HI9914controller Block diagram for an irrigation system with recirculation circuit...

Страница 20: ...ribed main tenance are not covered This warranty is limited to free of charge or replacement of the meter only if any malfunctioning is due to manufac turing defects If service is required contact the...

Страница 21: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax...

Отзывы: