Slope adjustment

• Rinse the electrode and the ground probe

thoroughly with water and immerse the
bottom 4 cm (1.5") in a pH10.01 (

HI 7010

)

or a pH4.01 (

HI 7004

) buffer solution.

• Stir the electrode and wait for the reading

to stabilize before adjusting the SLOPE trim-
mer to display the correct pH value at the
measured solution temperature; e.g. 4.01
(or 10.01) at 25°C (77°F). See the “pH
VALUES AT DIFFERENT TEMPERATURES” sec-
tion.

The pH calibration is now complete.

CONDUCTIVITY CALIBRATION

To calibrate the controller, first set it in measurement mode by pressing
the MEA button (the MEA LED lights up).

Make sure the conductivity probe has been properly connected and wired
to the controller and the meter is plugged to the mains.
The calibration should be performed at a temperature similar to that of
the liquid to be monitored.

Slope adjustment

• Pour in a beaker a solution of known conductivity value, which should

be close to the sample stream to be monitored. For example, if
measurements are in the 1.2 to 2.5 EC range, choose 

HI 7031

 (1.41

mS/cm @25°C). Similarly, if measurements are in the 4 to 6 mS/cm
range utilize 

HI 7031

 or 

HI 7039

 (5.00 mS/cm @25°C).

• Immerse the probe in the beaker ensuring that the holes on the probe

sleeve are completely covered.

CALIBRATION

Disconnect the pumps, the water nozzle and the alarm or assure that the
start of one of them will cause no damage.

ON, OFF or blinking LEDs have no effect on the measurement and
calibration of pH and EC.

pH CALIBRATION

To calibrate the controller, first set it in measurement mode by pressing
the MEA button (MEA LED is on).

Make sure the pH electrode and matching pin have been properly
connected and wired to the controller, and the meter is plugged to the
mains.
The calibration should be performed at a temperature similar to that of
the liquid to be monitored. Use the Hanna 

Checktemp

 (or other accu-

rate thermometer) as reference.
Remove the protective cap from the electrode.
During calibration, move the electrode and the ground probe (if in use)
together from one buffer to the next.

Offset adjustment

• Rinse the tip of the electrode with pH7.01

solution (

HI 7007

), dip the bottom 4 cm

(1.5") of the electrode and ground probe in
the pH7.01 buffer.

• Place also the 

Checktemp

 thermometer in

the solution.

• Wait for the reading to stabilize and adjust

the OFFSET trimmer to display the correct
pH value at the measured solution tem-
perature; e.g. 7.01 at 25°C (77°F). See the
“pH VALUES AT DIFFERENT TEMPERATURES”
section.

Содержание HI 9914

Страница 1: ...1 Instruction Manual HI 9914 Wall mounted Fertigation Controller w w w h a n n a i n s t c o m...

Страница 2: ...y on two large LCDs with backlight feature for a best easy reading TheECprobeisdesignedwithabuilt intemperaturesensorwhichallows the controller to automatically compensate for the temperature effect T...

Страница 3: ...erdosing of fertilizer or acid The controller is designed with relay outputs for pH and conductivity control 2A 240V each one corresponding to a status LED for visual check Three level sensors are use...

Страница 4: ...eeding pump output 20 Circulation pump output 21 Alarm output 1 LCD for Conductivity readings 2 EC SET button 3 EC slope adjustment trimmer 4 EC MEAsure button 5 Alarm LED 6 Reset button 7 LCD for pH...

Страница 5: ...djustable from 0 5 to 14 0 pH Calibration Manual 2 point with offset 2 pH and slope 80 to 120 trimmers Analogoutput 0 7V 5 0 5V pH Controller Output 2A 220V relay Other Features Timer Adjustable from...

Страница 6: ...erenttankvolumesandwithvariouscomplex ity of tank assembly For proper operation of the controller the tank assemblies must be set minding the following the pH electrode and EC probe should rapidly det...

Страница 7: ...less of the water composition or level The controllerwillrefillthetankifthewatergoesundernormallevelforabout 15 minutes or if the fill button is pressed Note The inlet of the circulation pump must be...

Страница 8: ...the NC pin when is on the COMMON pin is connected with the NO pin The NC pin is not accessible for the pH and EC pumps and the COMMON Place the low level sensor to indicate the minimum water level fo...

Страница 9: ...onnect one pin of the source to the COMMON pin of the alarm output Connect the alarm device between the NO output pin and the remaining source pin Acid and fertilizer dosing pump connection The pH and...

Страница 10: ...ke sure the controller is completely wired and all the connections for pumps alarm probes etc have been done Connect the mains power wires to the power supply input pins Replace the right panel lid an...

Страница 11: ...e conductivity and pH values are good for at least 30 seconds the controller goes to step 5 If the normal level sensor remains open for about 15 minutes the controller goes back to step 3 Step 5 The C...

Страница 12: ...is reached SETTINGTHEECCONTROL To adjust the Conductivity setpoint Push the SET button close to EC display to read the setpoint value ALARM CONDITIONS When an alarm condition is reached the ALARM LED...

Страница 13: ...swillbeactivefor 3 minutes and off for 12 minutes 3 12 15 minutes Setting the timer In order to activate the timed dosing feature put the TIMED CONT switch to the TIMED position To set the duty cycle...

Страница 14: ...25 C Immerse the probe in the beaker ensuring that the holes on the probe sleevearecompletelycovered CALIBRATION Disconnect the pumps the water nozzle and the alarm or assure that the start of one of...

Страница 15: ...trode stem or bulb If any scratches or cracks are present replace the electrode The connector must be perfectly clean and dry Cleaning procedure Soak the electrode in HI 7061 General Cleaning Solution...

Страница 16: ...perly or constantly reads zero or a value close to it Check the probe and cable for cracks and replace it if needed If the controller display shows 1 The cable may be shorted or the probe broken Repla...

Страница 17: ...ottle EC calibration solutions HI7031L 1 41 mS cm calibration solution 500 mL bottle HI7039L 5 00 mS cm calibration solution 500 mL bottle Maintenancesolutions HI70300L Electrode storage solution 500...

Страница 18: ...ystem utilized in nursery and cutting plants cultivation Reference Principitecnico agronomicidellafertirrigazioneedelfuorisuolo pag 115 publishedby VenetoAgricoltura CentroSperimentaleOrtofloricoloPod...

Страница 19: ...37 36 APPENDIX C INSTALLATIONEXAMPLES Block diagram for a typical installation of an irrigation system with HI9914controller Block diagram for an irrigation system with recirculation circuit...

Страница 20: ...ribed main tenance are not covered This warranty is limited to free of charge or replacement of the meter only if any malfunctioning is due to manufac turing defects If service is required contact the...

Страница 21: ...942 Japan Tel 03 3258 9565 Fax 03 3258 9567 Korea Tel 02 2278 5147 Fax 02 2264 1729 Malaysia Tel 603 5638 9940 Fax 603 5638 9829 Singapore Tel 6296 7118 Fax 6291 6906 SouthAfrica Tel 011 615 6076 Fax...

Отзывы: