11
HI 9110
HI 9210
HI 9210N
Range
pH
0.00 to 14.00 pH
°C
—
0.0 to 70.0°C
Resolution
pH
0.01 pH
°C
—
0.1°C
Accuracy
pH
±0.02 pH
(@20°C/68°F) °C
—
±0.5°C
Typical EMC
pH
±0.05 pH
±0.02 pH
Deviation
°C
—
±0.5°C
Calibration
Manual 2 points through trimmers
Offset Calibration
±1 pH
Slope Calibration
From 85 to 105% of nominal value
Temperature
Manual 0 to 100°C
Auto 0 to 70°C
Compensation
(32 to 212°F)
(32 to 158°F)
Electrode
FC 100B HI 1230B
HI 1217D
pH (
incl.
) pH (
incl.
) pH/°C (
incl.
)
Input Impedance
10
12
Ohm
Battery Type
4x1.5 volt, AA size (alkaline batteries);
Life
2000 hours of continuous use
Environment
0 to 50°C (32 to 122°F); 100% RH
Dimensions
196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
Weight
320 g (11.3 oz.)
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
SPECIFICATIONS HI 9110, HI 9210
& HI 9210N
& HI 9210N
& HI 9210N
& HI 9210N
& HI 9210N
30
If the batteries become weak:
-
HI 9023C and HI 9023CN
will display a
blinking additional decimal point.
-
HI 9024C and HI 9025C
will display "LOW
BAT".
-
HI 9110, HI 9210 and HI 9210N
will
display a "V" on the LCD. Note: HI 9110
HI 9110
HI 9110
HI 9110
HI 9110,
HI 9210
HI 9210
HI 9210
HI 9210
HI 9210 and HI 9210N
HI 9210N
HI 9210N
HI 9210N
HI 9210N are supplied with
the advanced "BEPS", Battery Error Preven-
tive System, that detects a low voltage condition in the battery.
The "V" symbol is to alert the user that the display will be shut-off
after about 5 hours of use. After that, the meter shuts down the
display, preventing the taking of erroneous measurements due to
low voltage.
Battery replacement must only take place in a
non hazardous area using the battery types speci-
fied in this instruction manual. To replace
rundown batteries, remove the two screws on
the rear cover of the instrument and replace all
four 1.5V AA batteries with new ones, while
paying attention to the correct polarity. Replace
the cover and tighten the two screws. The meter
is reset any time the batteries are removed. In
case of any functional problems, reset the meter
by removing and reinstalling the batteries.
pH
LOW BAT
V
BATTERY REPLACEMENT & AUTO-OFF
BATTERY REPLACEMENT & AUTO-OFF
BATTERY REPLACEMENT & AUTO-OFF
BATTERY REPLACEMENT & AUTO-OFF
BATTERY REPLACEMENT & AUTO-OFF