background image

20

• Once the electrode is submerged in the

buffer solution, the LCD will notify the user
if the reading is not stable by an intermit-
tent "NOT READY". Only when the reading
is stable, will it change to a blinking
"READY" and "CON".

• Press the CFM key to confirm the calibra-

tion. If the reading is not close to the
selected buffer, "WRONG 

1

BUF

" and

"WRONG  " will blink alternatively. If the
reading is close to the selected buffer, the
meter stores the reading (and adjusts the
offset point). The buffer value is then dis-
played on the primary LCD and the
secondary LCD will display another buffer
value (e.g."4.01").

• Press the CAL key and the calibration pro-

cess is ended with only the offset of the
meter calibrated. For best accuracy how-
ever, do not press CAL and proceed with a
two-point calibration.

TWO-POINT CALIBRATION

After calibrating at 6.86 or 7.01 pH (see above), for the second buffer
use pH 4.01 if you are going to measure acidic samples, and pH 9.18
or pH 10.01 if you are measuring alkaline solutions.
• Proceed as described in “One-point cali-

bration above but do not end the calibration
by pressing CAL.

• After the first calibration point is confirmed,

immerse the pH electrode into the second
buffer solution (pH 4.01, 9.18 or  10.01)
and stir gently.

Note: 

The electrode should be submerged

approximately 4 cm (1½") into the
solution with the temperature probe lo-
cated as close to it as possible.

• Select the second buffer value on the sec-

ondary display by pressing the 

Ï

°C 

or

Ð

°C

°C

°C

°C

°C key.

CAL

1

BUF

pH

pH

WRONG

WRONG

1

BUF

CFM

CAL

NOT READY

READY

CON

4 cm

(1½")

HI 7004

°C

°C

or

21

• When the “READY” and “CON” symbols

blink on the display, the reading is stable
and the calibration can be confirmed.

• Press the CFM key. If the reading is not close to the selected

buffer solution, “WRONG 

2

BUF

” and “WRONG  ” will blink alter-

natively. If the reading is close to the selected buffer, the slope
and the offset are calibrated. The values will be stored in memory
and the meter will return to the operating mode.

Note:

The meter will automatically skip the buffer that was used for
the first calibration to avoid erroneous calibration.  At least 1
full pH unit is required between the two buffers used for the
offset and slope calibration.  As a result, once calibrated at
either pH 6.86 or 7.01, the microprocessor will automatically
ignore the other one for the second point calibration. It will do
likewise for pH 9.18 and 10.01

Note:

During calibration, the secondary LCD displays the selected
buffer value. With HI 9025C, press the RANGE key to display
the buffer temperature during calibration.

PROCEDURE FOR HI 9110, HI 9210 & HI 9210N:

Manual Temperature Compensation
Note:

The pH/°C electrode of 

HI 9210N

 incorporates a temperature

sensor which performs automatic temperature compensation.
Consequently, the following 5 paragraphs are irrelevant to

HI 9210N

 and you can just proceed with pH calibration below.

• Press pH/°C SET key for 

HI 9110

 or the RANGE key for 

HI 9210

 to

select the manual temperature setting. The "°C" symbol will be
displayed to indicate the temperature setting mode.

• Remove the protective cap on the top of the meter to have access

to the calibration trimmers.

• Record the temperature of the buffer with a Checktemp C or an

READY

CON

WRONG

WRONG

2

BUF

CFM

pH

°C

SET

RANGE

°C

Содержание HI 9023C

Страница 1: ...nual HI 9023C HI 9023CN HI 9024C HI 9025C HI 9110 HI 9210 HI 9210N Portable Waterproof pH Meters Manufacturers since 1978 These Instruments are in Compliance with the CE Directives MANPHWPR3 09 04 PRI...

Страница 2: ...TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS 39 Recommendations for Users Before using these products make sure that they are entirely suitable for the environment in which...

Страница 3: ...ly to make sure that no damage has occurred during shipping If there is any noticeable damage notify your Dealer immediately Note Save all packing materials until you are sure that the instru ment fun...

Страница 4: ...rature It has only two buttons making it simple to use for even non technical personnel The calibration trimmers are sealed behind a protective cover so that they cannot be accidentally moved or readj...

Страница 5: ...efillable combination platinum ORP electrode Use titration HI 3230B Plastic body Ultem gel filled combination platinum ORP elec trode Use general purpose HI 4430B Plastic body Ultem gel filled combina...

Страница 6: ...012 Ohm Battery Type Life 4x1 5V AA size alkaline batteries approx 400 hours of continuous use auto shut off after 10 minutes of non use Environment 0 to 50 C 32 to 122 F 100 RH Dimensions 196x80x60 m...

Страница 7: ...N FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION H H H H HI I I I I 9024C H 9024C H 9024C H 9024C H 9024C HI I I I I 9025C 9025C 9025C 9025C 9025C 34 HI 2031B Glass body semimicr...

Страница 8: ...f continuous use auto shut off after 10 minutes of non use Environment 0 to 50 C 32 to 122 F 100 RH Dimensions meter 196x80x60 mm 7 7x3 1x2 4 kit 340x230x80 mm 13 8x9x3 1 Weight meter 425 g 15 oz kit...

Страница 9: ...L ELECTROLYTE SOLUTIONS IN FDA APPROVED BOTTLES HI 8071 3 5M KCl AgCl Electrolyte 4x50mL for single junc tion electrodes HI 8072 1M KNO3 Electrolyte 4x50 mL HI 8082 3 5M KCl Electrolyte 4x50 mL for do...

Страница 10: ...mL HI 7006L pH 6 86 Buffer Solution 460 mL HI 7007L pH 7 01 Buffer Solution 460 mL HI 7009L pH 9 18 Buffer Solution 460 mL HI 7010L pH 10 01 Buffer Sol 460 mL pH CALIBRATION SOLUTIONS IN FDA APPROVED...

Страница 11: ...isplay LOW BAT HI 9110 HI 9210 and HI 9210N will display a V on the LCD Note HI 9110 HI 9110 HI 9110 HI 9110 HI 9110 HI 9210 HI 9210 HI 9210 HI 9210 HI 9210 and HI 9210N HI 9210N HI 9210N HI 9210N HI...

Страница 12: ...s the ON OFF key OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE ON OFF 29 TAKING REDOX MEASUREMENTS HI 9023C HI 9025C only HI 9023C and HI 9025C have the cap...

Страница 13: ...ak in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately hour Removal of films dirt or deposits on the membrane junction Protein Soak in Hanna HI 7073 or HI 8073 Protein Cleaning Sol...

Страница 14: ...w during measure ments For AmpHel electrodes If the electrode does not respond to pH changes the battery is run down and the electrode should be replaced MEASUREMENT Rinse the electrode tip with disti...

Страница 15: ...abilize Note You can also calibrate the offset point of the HI 9210N tem perature sensor if needed For this purpose immerse the pH C electrode in a solution at a known temperature Then turn the 5 trim...

Страница 16: ...ves faulty readings The meter does not ac cept the 2nd buffer solution for calibration The reading drifts Display shows E1 or Display shows E2 or Display shows E3 or Display shows E4 or WRONG BUF 1 an...

Страница 17: ...instrument is now calibrated and will re main calibrated even when it is turned off Note The meter will lose the calibration if the batteries are removed CAL HI 7004 CFM 4 cm 1 24 Temperature has an e...

Страница 18: ...FM The E5 symbol will flash to indicate that the elec trode is still in the 7 01 solution Rinse the electrode with clean or distilled water first and then with pH 4 01 or 10 01 buffer Dip the electrod...

Страница 19: ...d 1 BUF indicators will be dis played The secondary LCD will display one of the 5 buffers e g 4 01 Press the C or C C C C C key until the secondary display changes to the buffer 6 86 CFM CAL 1 BUF pH...

Страница 20: ...EADY READY CON 4 cm 1 HI 7004 C C or 21 When the READY and CON symbols blink on the display the reading is stable and the calibration can be confirmed Press the CFM key If the reading is not close to...

Страница 21: ......

Отзывы: