2

PRELIMINARY EXAMINATION........................................................ 3
GENERAL DESCRIPTION ............................................................... 3
FUNCTIONAL DESCRIPTION HI 9023C & HI 9023CN ..................... 5
SPECIFICATIONS HI 9023C & HI 9023CN ..................................... 6
FUNCTIONAL DESCRIPTION HI 9024C & HI 9025C ........................ 7
SPECIFICATIONS HI 9024C & HI 9025C ....................................... 8
FUNCTIONAL DESCRIPTION HI 9110 ............................................ 9
FUNCTIONAL DESCRIPTION HI 9210 & HI 9210N ....................... 10
SPECIFICATIONS HI 9110, HI 9210 & 9210N ............................ 11
OPERATIONAL GUIDE ................................................................ 12
pH CALIBRATION....................................................................... 16
pH VALUES AT VARIOUS TEMPERATURES ................................... 24
TROUBLESHOOTING GUIDE ........................................................ 25
ELECTRODE CONDITIONING AND MAINTENANCE.......................... 26
TAKING REDOX MEASUREMENTS (HI 9023C & HI 9025C only) .... 29
BATTERY REPLACEMENT & AUTO-OFF ......................................... 30
ACCESSORIES ........................................................................... 31
ELECTRODE APPLICATION REFERENCE GUIDE .............................. 37
WARRANTY ............................................................................... 38
CE DECLARATION OF CONFORMITY ............................................. 39

Dear Customer,
Thank you for choosing a Hanna Product. Please read this instruction
manual carefully before using the instrument. It will provide you with
the necessary information for a correct use of the instrument, as well
as a precise idea of its versatility. If you require further technical
information, do not hesitate to e-mail us at [email protected].
These instruments are in compliance with the   directives EN 50081-1
and EN 50082-1.

TABLE OF CONTENTS

TABLE OF CONTENTS

TABLE OF CONTENTS

TABLE OF CONTENTS

TABLE OF CONTENTS

39

Recommendations for Users

Before using these products, make sure that they are entirely suitable for the environment in which they are used.
Operation of these instruments in residential area could cause unacceptable interference to radio and TV equipment,
requiring the operator to take all necessary steps to correct interference.
The glass bulb at the end of the electrode is sensitive to electrostatic discharges. Avoid touching this glass bulb at all
times.
During calibration of instruments, ESD wrist straps should be worn to avoid possible damage to the electrode by
electrostatic discharge.
To maintain the EMC performance of equipment, the recommended cables noted in the user's manual must be used.
In particular cases the instruments could turn off. In such cases, turn them on by pressing the ON/OFF key.
Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
To avoid electrical shock, do not use these instruments when voltages at the measurement surface exceed 24VAC or 60VDC.
To avoid damage or burns, do not perform any measurement in microwave ovens.

CE DECLARATION OF CONFORMITY

CE DECLARATION OF CONFORMITY

CE DECLARATION OF CONFORMITY

CE DECLARATION OF CONFORMITY

CE DECLARATION OF CONFORMITY

Содержание HI 9023C

Страница 1: ...nual HI 9023C HI 9023CN HI 9024C HI 9025C HI 9110 HI 9210 HI 9210N Portable Waterproof pH Meters Manufacturers since 1978 These Instruments are in Compliance with the CE Directives MANPHWPR3 09 04 PRI...

Страница 2: ...TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS TABLE OF CONTENTS 39 Recommendations for Users Before using these products make sure that they are entirely suitable for the environment in which...

Страница 3: ...ly to make sure that no damage has occurred during shipping If there is any noticeable damage notify your Dealer immediately Note Save all packing materials until you are sure that the instru ment fun...

Страница 4: ...rature It has only two buttons making it simple to use for even non technical personnel The calibration trimmers are sealed behind a protective cover so that they cannot be accidentally moved or readj...

Страница 5: ...efillable combination platinum ORP electrode Use titration HI 3230B Plastic body Ultem gel filled combination platinum ORP elec trode Use general purpose HI 4430B Plastic body Ultem gel filled combina...

Страница 6: ...012 Ohm Battery Type Life 4x1 5V AA size alkaline batteries approx 400 hours of continuous use auto shut off after 10 minutes of non use Environment 0 to 50 C 32 to 122 F 100 RH Dimensions 196x80x60 m...

Страница 7: ...N FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION FUNCTIONAL DESCRIPTION H H H H HI I I I I 9024C H 9024C H 9024C H 9024C H 9024C HI I I I I 9025C 9025C 9025C 9025C 9025C 34 HI 2031B Glass body semimicr...

Страница 8: ...f continuous use auto shut off after 10 minutes of non use Environment 0 to 50 C 32 to 122 F 100 RH Dimensions meter 196x80x60 mm 7 7x3 1x2 4 kit 340x230x80 mm 13 8x9x3 1 Weight meter 425 g 15 oz kit...

Страница 9: ...L ELECTROLYTE SOLUTIONS IN FDA APPROVED BOTTLES HI 8071 3 5M KCl AgCl Electrolyte 4x50mL for single junc tion electrodes HI 8072 1M KNO3 Electrolyte 4x50 mL HI 8082 3 5M KCl Electrolyte 4x50 mL for do...

Страница 10: ...mL HI 7006L pH 6 86 Buffer Solution 460 mL HI 7007L pH 7 01 Buffer Solution 460 mL HI 7009L pH 9 18 Buffer Solution 460 mL HI 7010L pH 10 01 Buffer Sol 460 mL pH CALIBRATION SOLUTIONS IN FDA APPROVED...

Страница 11: ...isplay LOW BAT HI 9110 HI 9210 and HI 9210N will display a V on the LCD Note HI 9110 HI 9110 HI 9110 HI 9110 HI 9110 HI 9210 HI 9210 HI 9210 HI 9210 HI 9210 and HI 9210N HI 9210N HI 9210N HI 9210N HI...

Страница 12: ...s the ON OFF key OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE OPERATIONAL GUIDE ON OFF 29 TAKING REDOX MEASUREMENTS HI 9023C HI 9025C only HI 9023C and HI 9025C have the cap...

Страница 13: ...ak in Hanna HI 7061 or HI 8061 General Cleaning Solution for approximately hour Removal of films dirt or deposits on the membrane junction Protein Soak in Hanna HI 7073 or HI 8073 Protein Cleaning Sol...

Страница 14: ...w during measure ments For AmpHel electrodes If the electrode does not respond to pH changes the battery is run down and the electrode should be replaced MEASUREMENT Rinse the electrode tip with disti...

Страница 15: ...abilize Note You can also calibrate the offset point of the HI 9210N tem perature sensor if needed For this purpose immerse the pH C electrode in a solution at a known temperature Then turn the 5 trim...

Страница 16: ...ves faulty readings The meter does not ac cept the 2nd buffer solution for calibration The reading drifts Display shows E1 or Display shows E2 or Display shows E3 or Display shows E4 or WRONG BUF 1 an...

Страница 17: ...instrument is now calibrated and will re main calibrated even when it is turned off Note The meter will lose the calibration if the batteries are removed CAL HI 7004 CFM 4 cm 1 24 Temperature has an e...

Страница 18: ...FM The E5 symbol will flash to indicate that the elec trode is still in the 7 01 solution Rinse the electrode with clean or distilled water first and then with pH 4 01 or 10 01 buffer Dip the electrod...

Страница 19: ...d 1 BUF indicators will be dis played The secondary LCD will display one of the 5 buffers e g 4 01 Press the C or C C C C C key until the secondary display changes to the buffer 6 86 CFM CAL 1 BUF pH...

Страница 20: ...EADY READY CON 4 cm 1 HI 7004 C C or 21 When the READY and CON symbols blink on the display the reading is stable and the calibration can be confirmed Press the CFM key If the reading is not close to...

Страница 21: ......

Отзывы: