100
101
pH
• Reinsert the cuvette into the instrument.
• Press the READ key to start the reading. The instrument displays
the pH value.
pH
pH
SPECIFICATIONS
Range
6.5 to 8.5 pH
Resolution
0.1 pH
Accuracy
±0.1 pH
Typical EMC
±0.1 pH
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow
to red tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93710-0
Phenol Red Indicator
5 drops
REAGENT SETS
HI 93710-01
Reagents for 100 pH tests
HI 93710-03
Reagents for 300 pH tests
For other accessories see page 128.
MEASUREMENT PROCEDURE
• Select the
pH method using the procedure described in the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press ZERO key. The display will show “-0.0-” when the meter
is zeroed and ready for measurement.
• Remove the cuvette and add 5 drops of HI 93710-0 Phenol
Red Indicator. Replace the cap and mix the solution.
10 mL
x 5