For HI 9214 and HI 931000: 

To take a pH

measurement simply submerge the tip (4cm/
1½") of the combination pH/temperature
electrode into the sample to be tested.
Turn the instrument ON. Select the pH mode
(

HI 8314

 and 

HI 8915

 only) or press the

RANGE key until the display changes to pH
(

HI 8424 

and 

HI 9622

 only).

Shake the electrode briefly and allow a couple
of minutes for the electrode to adjust and
stabilize. The display will show the pH value
compensated for temperature.

For HI 9214 and HI 931000:

After submerging the electrode into the sample
to be tested, wait until the stability indicator
"

" stops blinking. The pH readout is also

temperature compensated.
In order to take accurate pH measurements, make sure that the
instrument is calibrated for pH before use (see page 25).
For a faster response time when using 

refillable  electrodes

,

unscrew the refill hole cap.

If measurements are taken in different samples successively, it is
recommended to rinse clean the electrode thoroughly to eliminate
cross-contamination. After cleaning, it is recommended to rinse the
electrode with some of the sample to be measured.

TEMPERATURE  COMPENSATION

The pH reading is directly affected by temperature. In order for the
meter to measure the pH accurately temperature must be taken into
consideration.

HI 9214

 and 

HI 931000 

use an electrode with a built-in tempera-

ture sensor and therefore no additional temperature probe is needed.
To use the Automatic Temperature Compensation feature of 

HI 8314

,

HI 8424

,

 HI 8915 

and

 HI 9622

, submerge the temperature probe

into the sample as close to the electrode as possible and wait for a
couple of minutes. The displayed pH reading is compensated for the
temperature variance associated with the effects of temperature upon
the electrode.
If you know the temperature of the sample you can also manually

4 cm

(1

½

")

RANGE

pH

For HI 931000:

Attach the pH/°C electrode to the DIN socket on the top of the meter,
tighten the retainer ring and turn the meter on by pressing ON/OFF.

TAKING  pH  MEASUREMENTS

For HI 8014:

To take a pH measurement simply submerge
the tip (4cm/1½") of the electrode and a
ChecktempC (or an accurate thermometer) into
the sample to be tested.
Turn the rotary knob to the °C position to
display temperature setting on the LCD.
Measure the temperature of the sample with a
ChecktempC (or an accurate thermometer), e.g.
25°C.
Using the provided screwdriver, turn the temperature trimmer to
display the measured temperature value.

Turn the rotary knob to the pH position to
display the pH measurement.
Shake the electrode briefly and allow a couple
of minutes for the electrode to adjust and
stabilize. The display will show the pH value
compensated for temperature.

For HI 8314, HI 8424, HI 8915, HI 9214, HI 931000 and
HI 9622:

To take a pH measurement simply submerge
the tip (4cm/1½") of the electrode and the
temperature probe into the sample to be
tested.

ins rum n s

CFM

°C

CAL

OFF

ON

pH

HI 931000

ON

OFF

°C

4 cm

(1

½

")

OFF

mV

°C

pH

OFF

mV

°C

pH

pH

4 cm

(1

½

")

°C

ºC

Содержание HI 8014

Страница 1: ...Instruction Manual HI 8014 HI 8314 HI 8424 HI 8915 HI 9214 HI 931000 HI 9622 Portable pH Meters h t t p w w w h a n n a i n s t c o m These Instruments are in Compliance with the CE Directives...

Страница 2: ...panel HI 8314 comes supplied with HI 1230B combination double junction gel pH electrode HI 7669AW temperature probe a 9 volt battery and a calibration screwdriver HI 8424 is a portable microprocessor...

Страница 3: ...HI 8915 comes supplied with HI 1230B combination double junction gel pH electrode HI 7669AW temperature probe a 9 volt battery and a calibration screwdriver HI 9214 is a compact microprocessor based...

Страница 4: ...4 9 6 pH ON OFF mV C pH HI 8314 membrane pHmeter STD 1nd Calibration at pH 6 86 or 7 01 SLOPE 2nd Calibration at pH 4 01 or 10 01 CALIBRATION HI 8014 Range pH 0 00 to 14 00 mV 0 to 1999 Resolution pH...

Страница 5: ...ive connection pH ON OFF HOLD mV C pH HI 8915 ATC pHmeter mV OUTPUT SLOPE 4 01 10 01 OFFSET 6 86 7 00 CALIBRATION 1 7 2 9 4 10 5 11 8 3 6 12 SPECIFICATIONS HI 8314 SPECIFICATIONS HI 8314 SPECIFICATION...

Страница 6: ...0 1 Accuracy pH 0 01 20 C 68 F mV 1 C 0 5 Typical EMC pH 0 02 Deviation mV 1 C 0 8 pH Calibration Manual 2 point through offset and slope trimmers Offset Calibration 1 pH Slope Calibration from 85 to...

Страница 7: ...irm calibration data HI 9214 Range pH 0 00 to 14 00 C 0 0 to 70 0 Resolution pH 0 01 C 0 1 Accuracy pH 0 02 20 C 68 F C 0 5 Typical EMC pH 0 08 Deviation C 1 pH Calibration Automatic 2 point with 3 me...

Страница 8: ...00 ISE 0 0 to 399 9 mV ORP 400 to 1999 mV C 0 0 to 100 0 Resolution pH 0 01 ISE 0 1mV ORP 1mV C 0 1 Accuracy pH 0 01 20 C 68 F ISE 0 2mV ORP 1mV C 0 4 Typical EMC pH 0 02 Deviation ISE 0 2 ORP 1 C 0...

Страница 9: ...l compensation when the temperature probe is disconnected 9 CFM to confirm calibration data HI 931000 Range pH 0 00 to 14 00 C 0 0 to 70 0 Resolution pH 0 01 C 0 1 Accuracy pH 0 01 20 C 68 F C 0 5 Typ...

Страница 10: ...9622 the display will show the previ ous calibration date month date for 3 seconds For HI 9214 Connect the pH electrode to the screw type connector on the meter Be sure that the electrode and the met...

Страница 11: ...ional temperature probe is needed To use the Automatic Temperature Compensation feature of HI 8314 HI 8424 HI 8915 and HI 9622 submerge the temperature probe into the sample as close to the electrode...

Страница 12: ...E C compensate for it HI 8424 and HI 9622 only In this case the temperature probe must be disconnected With HI 8314 and HI 8915 when the temperature probe is discon nected the pH reading will be compe...

Страница 13: ...HI 8014 HI 8314 and HI 8915 to NBS standards use pH 6 86 HI 7006 or HI 8006 instead of pH 7 01 and pH 9 18 HI 7009 or HI 8009 instead of pH 10 01 PROCEDURE FOR HI 8014 Remove the protective cap from...

Страница 14: ...st the SLOPE trimmer on the lower right of the front panel until the LCD shows the pH value at the noted temperature The pH calibration is now complete HI 7007 4 cm 1 C C pH HI 7004 OFFSET pH SLOPE pH...

Страница 15: ...on the previous page HI 7004 pH pH CFM RANGE C HI 7007 C PROCEDURE FOR HI 8424 Switch the meter on after connecting the pH electrode and the temperature probe Remove the protective cap from the elec t...

Страница 16: ...first buffer solution ins e rum n s ON OF F CA L CF M HI 921 4 stic k pHm eter HI 7004 ON OFF CAL CFM HI 7007 4 cm 1 CFM PROCEDURE FOR HI 9214 Turn the meter on Note the stability indicator blinks wh...

Страница 17: ...offset one point calibration For best accuracy however it is recommended that a two point calibra tion is performed The LCD will now blink 4 01 pH 10 01 can be selected by pressing the CAL key Rinse w...

Страница 18: ...03 or 10 06 If the buffer temperature is 50 C the display should show pH 4 06 or 6 98 or 9 82 pH CAL The meter will request pH 4 01 or 10 01 buffer showing Er 4 This indicates that the instrument is...

Страница 19: ...libration An out of order pH electrode The electrode is not working or the refer ence junction is clogged Erroneous buffer so lution used for slope calibration An out of order pH electrode A clogged o...

Страница 20: ...ffer so lution used for slope calibration An out of order pH electrode Out of range pH scale No temperature probe attached Out of range mV scale Erroneous pH 7 buffer solution used for off set calibra...

Страница 21: ...ish the metal tip with a lightly abrasive paper paying attention not to scratch the surface and wash thoroughly with water If the bulb and or junction are dry soak the electrode in HI 70300 Storage So...

Страница 22: ...ention to its polarity Note the instrument should be recalibrated after changing the battery HI 8114 HI 8414 HI 9214 9 VOLT BATTERY SLIDE HI 9214 HI 8014 HI 8314 HI 8424 HI 8915 HI 9622 HI 931000 HI 8...

Страница 23: ...ning Solution 460 mL HI 7073M Protein Cleaning Solution 230 mL HI 7073L Protein Cleaning Solution 460 mL DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE DISPLAY CODE GUIDE...

Страница 24: ...ution 230 mL HI 7077L Oil Fat Cleaning Solution 460 mL ELECTRODE CLEANING SOLUTIONS IN FDA APPROVED BOTTLE HI 8061M General Cleaning Solution 230 mL HI 8061L General Cleaning Solution 460 mL HI 8073M...

Страница 25: ...4 7 12 mm 0 5 FC 210 120 mm 4 7 12 mm 0 5 9 5mm DIA 0 37 FC 220 110 mm 4 3 12 mm 0 5 FC 911 HI 1330B HI 1310S Glass body semimicro single junction refillable combination pH electrode Use laboratory H...

Страница 26: ...931001 pH and ORP electrode simulator with LCD and 1 m 3 3 coaxial cable ending in female BNC connectors HI 710007 Shockproof blue rubberboot for HI 931000 formerly HI 935BTB HI 710008 Shockproof oran...

Страница 27: ...read HI 2031B FC 200B HI2020S FC 200S 5 Cheese FC 200B FC200S 6 Dairyproducts FC 911B FC 100B 7 Dirtywater HI 1910B HI 1912B 8 Emulsions HI 1053B HI1050S 9 Environment HI 1230B HI1210S 10 Flasks HI 13...

Страница 28: ...netic fields it is recommended to keep parallel the cables of the two probes and as close as possible to the instrument In particular cases HI 8915 could change its operating mode In such cases press...

Страница 29: ...ITERATURE HANNA LITERATURE LAB RECORDING WATER ANALYSIS HANDBOOK ENVIROCARE GENERAL CATALOG These and many others catalogs handbooks and leaflets are avail able from Hanna To receive your free copy co...

Отзывы: